• Title/Summary/Keyword: tetracyanonickelate(II)

Search Result 1, Processing Time 0.014 seconds

A Continuous Monitoring System for the Determination of Cyanide Ions Using Nickel Complexation Reaction (흡수분광법을 적용한 시안화이온의 연속 모니터링 장치 구성 및 적용)

  • Kim, Kyung Min;Lee, Sang Hak;Kim, Young Ho;Kim, Gyu Man;Oh, Sang-Hyub
    • Applied Chemistry
    • /
    • v.16 no.1
    • /
    • pp.25-28
    • /
    • 2012
  • A determination method of cyanide ion (CN-) using nickel complexation reaction by continuous monitoring system. The mechanical parameters and chemical conditions of the complexation reaction were investigated prior to application of continuous monitoring system for determination of cyanide. On the optimized conditions, the calibration curve was linear over the range from 5.0×10-6 to 1.0×10-4 M. In this range, 2.40% of the reproducibility (RSD, n=3) was obtained. The limit of detection (3σ/s) was calculated to be 1.8×10-6 M.