• Title/Summary/Keyword: stereotactic rediosurgery

Search Result 1, Processing Time 0.015 seconds

Assessment of Imaging Distortion in Magnetic Resonance Imaging for Stereotactic Radiosurgery: Through Phantom Study (뇌정위 방사선수술 시스템을 위한 자기공명영상의 공간적 왜곡의 측정 : 모형실험을 통한 연구)

  • 박선원;한문희;김동규;정현태;송인찬
    • Investigative Magnetic Resonance Imaging
    • /
    • v.4 no.1
    • /
    • pp.7-13
    • /
    • 2000
  • Purpose : To assess the distortion of MRI with the Leksell stereotactic radiosurgery system in variable pulse sequence and imaging plane through phantom study, to find most adequate imaging plane and pulse sequence for stereotactic radiosurgery system. Materials and methods : We made the phantoms for MRI and get images in variable conditions and analyzed the image distortion using image analysis program, and statistically using paired student t-test. Results : The transeverse plane images had acceptable error ranges bless than 1.5mm) in all pulse sequence in both the analysis of fiducial marker in stereotactic G-frame and the phantom study. The coronal plane images had unacceptable large errors (more than 1.7mm) in the analysis of fiducial marker in the stereotactic G-frame, but had corrected small errors (less than 1.5mm) in the phantom study. Conclusion : We find from the phantom study that the present MR machines are adequate for stereotactic surgery system in frequently used pulse sequences, and imaging planes.