• Title/Summary/Keyword: reductive vapor generation ICP-AES

Search Result 1, Processing Time 0.013 seconds

Simultaneous Determination of Mercury and Arsenic by Reductive Vapor Generation-ICP-AES (환원 증기 발생법-유도결합 플라즈마 원자방출 분광계를 이용한 수은과 비소의 동시 분석법)

  • Shin, Hyung-Seon;Choi, Man-Sik;Kim, Kang-Jin
    • Analytical Science and Technology
    • /
    • v.12 no.4
    • /
    • pp.273-278
    • /
    • 1999
  • Simultaneous determination of mercury and arsenic has been studied by reductive vapor generation-ICP-AES. Samples were digested with a microwave digestion system and extracted with acids. Reductive vapor generation was carried out with 6N HCI and 2% $NaBH_4$. Detection limit of Hg and As are found to be 0.06 ppb and 0.08 ppb, respectively. Measured values of Hg and As in inorganic samples agree well with reference value of SRMs, but the recovery of As from organic samples is obtained approximately 80% of the reference values.

  • PDF