• Title/Summary/Keyword: nonmagnetic metal screw

Search Result 1, Processing Time 0.015 seconds

The usability of the image according to the frequency encoding gradient direction conversion in fixation using the non magnetic metal screw (비 자성 금속 screw를 이용한 고정술에서 주파수 부호화 경사 방향 변환에 따른 영상의 유용성)

  • Cho, Jae-Hwan;Lee, Hae-Kag;Park, Cheol-So
    • Journal of Digital Contents Society
    • /
    • v.12 no.1
    • /
    • pp.49-55
    • /
    • 2011
  • Because of causing the geometrical transformation of the magnetic field, the patient implementing the fixation using the nonmagnetic metal screw causes the magnetic susceptibility artifact at an image. Thus, in this research, the distortion measure of the image according to the frequency oblique direction conversion tried to be compared in the magnetic susceptibility artifact occurence. First, the itself phantom inserting the nonmagnetic metal screw of the titanium component was made and the region of interest was set up and the frequency oblique direction the anterior - back side was converted to the right-to-left direction in the axial image and a right-to-left was converted to the upper side - bottom side in the coronal plane and the upper - bottom side was converted to the anterior - back side in the sagittal plane and the distortion measure of the region of interest was compared, it observed. In a result, when converting the frequency oblique direction, the distortion difference of the region of interest could be confirmed and it is considered to enhance the diagnostics efficiency changing the oblique direction appropriately.