• Title/Summary/Keyword: miniatured transducer

Search Result 2, Processing Time 0.017 seconds

Miniaturization of disposable functional flow tube (기능성 일회용 호흡관의 소형화 연구)

  • Kim, Kyung-Ah;Lee, Tae-Soo;Cha, Eun-Jong
    • Journal of Sensor Science and Technology
    • /
    • v.14 no.4
    • /
    • pp.250-257
    • /
    • 2005
  • Respiratory tubes with the length of 35 mm and the diameter of 10, 15, and 20 mm were made and both the static($P_{S}$) and dynamic($P_{D}$) pressures were measured for steady flow rates ranging 1-12 l/sec. Regression analysis resulted successful fitting of $P_{S}$ and $P_{D}$ data with quadratic equations with correlation coefficients higher than 0.99. The measurement standards of the American Thoracic Society (ATS) were applied to $P_{S}$ data, which demonstrated the smallest tube diameter of 15 mm to satisfy the ATS standards. The maximum $P_{D}$ value of the velocity type transducer with the diameter of 15 mm was estimated to be 75 cm$H_{2}O$, implying approximately 7 times larger sensitivity than the widely used pneumotachometer. These results showed that the velocity type respiratory air flow transducer is a unique device accomplishing miniaturization with the sensitivity increased, thus would be of great advantage to develop portable devices.

Design Criteria to Miniaturize the Single Use Functional Respiratory Air Flow Tube

  • Kim Kyung Ah;Lee Tae Soo;Cha Eun Jong
    • Journal of Biomedical Engineering Research
    • /
    • v.26 no.3
    • /
    • pp.163-169
    • /
    • 2005
  • Respiratory tubes with a length of 35mm and diameters of 10, 15, and 20mm were made for experimental purpose, and both the static$(P_s)\;and\;dynamic(P_D)$ pressures were simultaneously measured for steady flow rates ranging 1-12//sec. Least squares analysis resulted successful fitting of $P_s\;and\;P_D$ data with quadratic equations with correlation coefficients higher than 0.99(P<0.0001). The spirometric measurement standards of the American Thoracic Society(ATS) were applied to $P_s$ data, which demonstrated the smallest tube diameter of 15mm to satisfy the ATS standards. The maximum $P_D$ value of the velocity type transducer(the functional single use respiratory air flow tube) with the diameter of 15mm was estimated to be approximately $75cmH_2O$, implying more than 7 times larger sensitivity than the widely used pneumotachometers. These results showed that the velocity type respiratory air flow transducer is a unique device accomplishing miniaturization with the sensitivity increased, thus would be of great advantage to develop portable medical devices.