• Title/Summary/Keyword: elemental mercury cold vapor

Search Result 1, Processing Time 0.015 seconds

Determination of Inorganic Mercury in Urine and Airborne by Cold Vapor Atomic Absorption Spectrophotometry (냉각증기 원자흡수분광법에 의한 뇨 및 공기중의 무기수은 정량에 관한 연구)

  • Oh, Doe Suk;Oh, Se Min
    • Journal of Korean Society of Occupational and Environmental Hygiene
    • /
    • v.1 no.2
    • /
    • pp.136-143
    • /
    • 1991
  • Inorganic mercury in urine and airborne was determined by cold vapor atomic absorption spectrophotometry. Detailed sampling methods and analylical results are as follows : 1. 100~200ml of urine for each person was taken in 250 ml borosilicate bottle and $K_2S_2O_8$ (0.1g/100ml urine) was added to prevent bacterial contamination. About 1001 air of workingplace was absorbed in l0ml of absorbing solution. Urine samples and absorbing solution tubes were stored at $4^{\circ}C$. Dillution solution to prepare standard solution used deionized water (D.W) for urine and absorbing solution (A.S) for air. 2. 1n this procedure deteclion limit was 1ng/ml and mercury contents of blank reagent solution was 1~2ng/ml. 3. Calibration range was $0.02{\sim}0.1{\mu}g/ml$ and in this range r.s.d for each calibration curve in D.W and A.S and ${\pm}7.9%$ and ${\pm}3.7%$, respectively. 4. Repeatability (n=5 times, conc. $0.05{\mu}g/ml$) was ${\pm}5.8%$, in D.W. and ${\pm}4.4%$ in A.S, respectively. 5. Recovery for urine adding spiked concentration ($0.05{\mu}g/ml$) was about 90%. 6. Analytical result of samples was $1{\sim}139{\mu}g/l$ in urine and ${\sim}0.127mg/m^3$ in airborne.

  • PDF