NMHCs measurement using a cryogenic concentration system and application on gas samples (저온농축장치를 이용한 NMHCs의 측정법과 가스 시료에 적용)

  • Kim, Su Ha;Moon, Dong Min;Kim, Jin Seog
    • Analytical Science and Technology
    • /
    • v.25 no.6
    • /
    • pp.375-381
    • /
    • 2012
  • In this study, we have developed a cryogenic concentration system for the analysis of non-methane hydrocarbons (NMHCs) in air sample. NMHCs with nmol/mol level of concentrations in the atmosphere were quantified by a comparative method with certified reference gas mixture. GC/FID with cryogenic concentration system operated in the range from $-67^{\circ}C$ to $180^{\circ}C$ was adopted. After cryogenic concentration, a fast desorption by heating could achieve a large amount of sample injection into the GC/FID. The linearity of the system was verified with the peak areas of NMHCs of the concentration amount of sample. The reproducibility is less than 10% and the limit of detection (LOD) is 0.1 nmol/mol. For the application of this system, we analyzed NMHCs in gas samples taken from air and soil in Daejeon, Seosan and Goheung.