• Title/Summary/Keyword: Wii balance board (WBB)

Search Result 11, Processing Time 0.015 seconds

Reliability and validity of pelvic mobility measurement using a cushion sensor in healthy adults

  • Jung, Seung-Hwa;Kim, Su-Kyeong;Lee, Ji-Hyun;Choi, Soo-Ih;Park, Dae-Sung
    • Physical Therapy Rehabilitation Science
    • /
    • v.9 no.2
    • /
    • pp.74-81
    • /
    • 2020
  • Objective: To prevent low back pain, an objective evaluation tool to evaluate pelvic mobility and exercise to improve the flexibility of the lumbar region is needed. The purpose of this study was to compare the results of pelvic mobility measurements using the Wii Balance Board (WBB) and Sensbalance Therapy Cushion (STC), evaluate the usefulness of the STC as a tool for measuring pelvic mobility. Design: Cross-sectional study. Methods: Fifty healthy subjects participated in this study. The subjects performed pelvic mobility range, proprioception, reaction time and reach of the arm using the STC. The pelvic movement parameter was measured two times to determine the intra-rater reliability. To measure the correlation between lumbar muscle tension and pelvic mobility, Myovision was used to measure tension of L4, L5 level erector spinae muscle. Correlations between measured variables were checked to determine the validity of the pelvic mobility assessment tool. Results: STC showed high test-retest reliability in pelvic tilt measurement and reaching task [intraclass correlation coefficients (3,1)=0.804-0.915]. The relationship between WBB and STC showed a significant positive correlation with the pelvic tilt and reaching task (p<0.05). Posterior tilt and erector spinae activation (Lt. L5) showed a significant negative correlation (p<0.05). Left, right tilt and erector spinae activation (L5) showed a significant negative correlation (p<0.05). Conclusions: This study confirmed the advantages of the STC and found efficiency as an objective measuring device of pelvic mobility.