• Title/Summary/Keyword: Weighting value

Search Result 351, Processing Time 0.034 seconds

Evaluation of Organ and Effective Dose using A PC-Based Monte Carlo Program in AEC Mode and Fix Mode for the whole spine antero-posterior radiography (전 척추 전.후 방향 검사 시 AEC Mode와 Fix Mode에서 PC-Based Monte Carlo Program을 이용한 장기선량 및 유효선량 평가)

  • Kim, Jeong Jin;Jang, Seong Won;Park, Jang Heum;Lee, Kwan Seob;Ha, Dong Yoon
    • Korean Journal of Digital Imaging in Medicine
    • /
    • v.14 no.2
    • /
    • pp.23-31
    • /
    • 2012
  • There are AEC mode and fix mode to exposure when the whole spine antero-posterior radiography is done by using DR equipment. This study compared the utility of fix mode to AEC mode, by evaluating organ dose and effective dose and by examining the quality of radiographic image. GE DEFINIUM 8000 and ART-200X Rando Phantom manufactured by Flukebiometical were used for this study. The Rando phantom was set in front of wall detector of X-rays equipment. AEC mode was set at 80kVp and Fix mode was set at 80kVp, 25mAs, 32mAs, 40mAs, and 50mAs. Whole spine AP image were aquired by combining C, T-L and L-S spine images obtained through 3 exposures. When obtaining C, T-L and L-S spine images, were checked for Air kerma (mGy) value calculated by UNFORS Xi meter attached at the phantom surface of center of radiation field. The effective and organ doses were compared by PCXMC program (PC-Based Monte Carlo Program). The quality of obtained radiographic image was evaluated visually by 3 radiologists using resolution chart. When the effective doses was calculated based on tissue weighting factor of ICRP-103, 1.278mSv was measured by AEC mode, and Fix mode measured 0.405mSv at 25mAs, 0.518mSv at 32mAs, 0.649mSv at 40mAs, and 0.810mSv at 50mAS. In addition, the organ dose measured with esposure at 25mAs by Fix mode was almost equivalent to the organ dose by AEC mode, at the esophagus, thyroid, oral mucosa, salivaly glands located at the cervical spine part, while the organ dose by Fix mode was in general lower than the organ dose by AEC mode at the other organs. When Fix mode at 32mAs, 40mAs, and 50mAs was compared to AEC mode for organ dose in 26 organs, AEC mode had higher measurement in 21 organs but not for than brain, trachea, thyroid, oral mucosa, and salivaly glands which are located at the cervical spine part. The image quality evaluated by resolution test chart was much higher with AEC mode than the quality with Fix mode at all exposure conditions. However, while the image quality of cervical spine exposured at 50mAs by Fix mode was lower than the quality of AEC mode, thoraco-lumbar spine and lumbo-sacral spine were calculated and the quality was similar to AEC mode. Scoliosis occurs mainly at thoraco-lumbar and lumbo-sacral spine, not at cervical spine. Compared to AEC mode, Using the appropriate protocol (80kVp, 50mAs) of fix mode for whole spine AP radiography was thought to be useful because the image quality of the thoraco-lumar and lumbo-sacral spine was similar on AEC mode, Also organ and effective doses can be decreased with Fix mode. Therefore, It is considered that fix mode can be used properly with AEC mode for whole spine AP radiography when considering patient's body posture.

  • PDF