Accurate determination of minor isotope ratios in individual plutonium-uranium mixed particles by thermal ionization mass spectrometry

  • Lee, Chi-Gyu;Park, Jinkyu;Lim, Sang Ho
    • Nuclear Engineering and Technology
    • /
    • v.50 no.1
    • /
    • pp.140-144
    • /
    • 2018
  • Isotopic analyses of plutonium and low-enriched uranium mixtures with particle sizes of $0.6-3.3{\mu}m$ were performed using thermal ionization mass spectrometry with a continuous heating method to verify its effectiveness for the accurate analysis of minor isotopes without sample pretreatment. The mixed particles used in this study were prepared from a mixed solution of plutonium (SRM 947) and uranium (U010, $^{235}U$ 1% enriched) reference materials. The isotope ratios for plutonium in the individual mixed particles, including $^{238}Pu/^{239}Pu$, $^{241}Pu/^{239}Pu$ as well as $^{240}Pu/^{239}Pu$, and $^{242}Pu/^{239}Pu$, were in good agreement with the certified values despite the isobaric interference of $^{238}U$ and $^{241}Am$. The isotope ratios for uranium in the mixed particles also agreed well with the certified values within the range of error. However, the isotope ratios for minor isotopes, such as $^{234}U$ and $^{236}U$, in the particles with diameters of less than approximately $1.8{\mu}m$ could not be measured because numbers of $^{234}U$ and $^{236}U$ atoms in analyzed particles are too low. These results indicate that thermal ionization mass spectrometry with a continuous heating method is applicable for the analysis of trace amounts of plutonium isotopes, including $^{238}Pu$ and $^{241}Pu$, despite the presence of the respective isobars $^{238}U$ and $^{241}Am$ in the microsamples.