• Title/Summary/Keyword: Tri-coil bellows transducer

Search Result 2, Processing Time 0.016 seconds

A Study on Frequency Characteristics According to the Output Transmission Method of Round Window Driving Middle Ear Implants (정원창 구동형 인공중이의 출력 전달 방식에 따른 주파수 특성 연구)

  • Seong, KiWoong;Shin, DongHo;Na, SungDae;Lee, JyungHyun;Kim, MyoungNam;Cho, Jin-Ho
    • Journal of Biomedical Engineering Research
    • /
    • v.39 no.5
    • /
    • pp.183-187
    • /
    • 2018
  • In this paper, we investigated the efficiency according to the output transmission method of the round window driving type AMEIs (active middle ear implants) through the cadaveric experiment. For the experiment, we fabricated DRT (direct rod transducer) and FMT (floating mass transducer) type vibrational transducers based on our previous studies and conducted their output characteristics were measured. TCBT (tri-coil bellows transducer) and DFMT (differential floating mass transducer) were implemented with the same driving force and electrical characteristics as one of DRT and FMT, respectively. In the experiment using three human temporal bone, normal stapes vibration was measured with 1 Pa in front of tympanic membrane, and then was compared with each output of transducers. From the comparison, the DRT type vibration transducer was superior in overall energy transfer efficiency, especially in the low frequency range. There was no difference in implantation difficulty between the two transducers. The results of this study suggest that the DRT type vibrational transducer is more efficient and needs further study to overcome the low frequency degradation in round window approaching with FMT.

Design of Signal Processing Circuit for Semi-implantable Middle Ear Hearing Device with Bellows Transducer (벨로즈형 진동체를 갖는 반이식형 인공중이용 신호처리회로 설계)

  • Kim, Jong Hoon;Shin, Dong Ho;Seong, Ki Woong;Cho, Jin-Ho
    • Journal of rehabilitation welfare engineering & assistive technology
    • /
    • v.11 no.1
    • /
    • pp.63-71
    • /
    • 2017
  • In this paper, a signal processing circuit for semi-implantable middle ear hearing device is designed using the TCBT which is recently proposed for a new middle ear transducer that can be implanted at round window of cochlea. The designed semi-implantable hearing device transmits digital sound signal from external device located at behind the ear to the internal device implanted under the skin using inductive coupling link methods with high efficiency. The coils and signal processing circuits are designed and implemented considering the total transmission and reception distance including skin thickness of temporal bone for the semi-implantable hearing device. And also, to improve the data transmission efficiency, the output circuits which can supply sufficient signal power is designed. In order to confirm operation of semi-implantable hearing device using inductive coupling link, the circuit analysis was performed using PSpice, and the performance was verified by implementing a signal processing board of an available size.