Dose Assessment According to Application of Carbon Fabric Blanket During Radiation Therapy of the Spine Metastasis Cancer (척추 전이암 환자의 방사선치료 시 Carbon Fabric Blanket 적용에 따른 선량평가)

  • Yang, Myung-Sic;Kim, Jung-Soo;Lee, Sun-Young;Kwon, Hyoung-Cheol
    • Journal of radiological science and technology
    • /
    • v.42 no.1
    • /
    • pp.61-66
    • /
    • 2019
  • The purpose of this study was to improve the unstable treatment posture by placing the Carbon fabric blanket on the couch which was used for the patient fixation for the unstable posture from the severe pain caused by the neuromuscular pressure of the spinal metastatic cancer patient and to analyze the dose difference caused by the energy loss of high energy radiation. Using a linear accelerator, a FC-65G was installed at a depth of 5 cm at a solid phantom at 6 MV and 10 MV energies. The SAD was 100 cm, Gantry angle was $0^{\circ}$, a Cotton and Carbon blanket with a thickness of 1 cm on the couch, The blankets were placed on the couch and the dose was measured according to field size. For the dose measurement, and the dose was measured at 100 MU each time, and the mean value was calculated by repeating the measurement three times in order to reduce the error. The results showed that the difference rate in dose between Carbon blanket and Cotton blanket was respectively -0.54% and -0.75% based on the absence of the blanket(Non). Therefore, it is considered that the use of Carbon fabric blanket, which reduces the patient's pain and does not affect the depth dose, may be useful during radiation therapy of the spine metastasis cancer.

Geometrical Limitations in Non-coplanar Treatment (비동일 평면 치료에서의 기하학적인 제약)

  • 이병용
    • Progress in Medical Physics
    • /
    • v.8 no.1
    • /
    • pp.31-35
    • /
    • 1997
  • Purpose: To analyse the geometrical limitations in non-coplanar treatment for applying this result to treatment planning. Material and Method: The ranges of gantry movement were mesured for the treatment sites with or without EPID and various couch angle. Resultas and conclusitn : The gantry range of motions for various situations were quantitized for applying these results to treatment planning.

  • PDF

Influence of Aquaplastic Mask Immobilization on Patient Positioning in Head & Neck Radiotherapy (두경부암 방사선 치료 시 Aquaplastic Mask가 환자 고정에 미치는 영향)

  • Lee, Ho-Yeon;Bae, Sun-Myung;Kim, Jin-San;Kang, Tae-Young;Kwak, Jung-Won;Back, Geum-Mun
    • The Journal of Korean Society for Radiation Therapy
    • /
    • v.24 no.2
    • /
    • pp.175-181
    • /
    • 2012
  • Purpose: This study aimed to analyze the influence of aquaplastic mask which is used for the radiotherapy for patients with head and neck cancer on patient positioning by analyzing the results of 2D/2D matching with the use of Exactrac 6D couch system. Materials and Methods: 2D/2D matching-based Image guidance was performed with the use of Exactrac 6D couch system on a total of 14 patients with head and neck cancer who received radiotherapy at this hospital in a period from June 2010 to June 2011. The adjusted 6D values obtained after the image guidance were classified into values related with patients and values not related with patients as to the correlation between the results. Results: In the analysis of the relation of each adjusted value to patients, the coefficient of determination ($R^2$) of longitudinal and pitch direction was 0.76, of longitudinal and yaw direction was 0.48, and of pitch and yaw direction was 0.42 although the coefficient of determination for above factors was 0.03 or less when the dependency of patients was excluded. Conclusion: It was demonstrated that intrinsic factors of aquaplastic mask had influence on the immobilization of patients by aquaplastic mask although various factors related with patients had influence on the immobilization of patients. It is believed that if these factors are adjusted, the treatment of patients with head and neck cancer can be more accurate.

  • PDF

Practicability Assessment of Spherical Type Mechanical Check Device (SMCD) (Mechanical Check용 Spherical Device의 제작 및 특성 평가)

  • Lee, Byung-Koo;Kim, Gun-Oh;Kweon, Young-Ho
    • The Journal of Korean Society for Radiation Therapy
    • /
    • v.19 no.1
    • /
    • pp.55-62
    • /
    • 2007
  • Purpose: Digital medical image commenced with an introduction of PACS has become more popular today in the radiation diagnosis and medical treatment and made great progress, in particular, for medical testing field, whereas it has made slow progress for radiation therapy area. In order to accommodate the current trend of digital from analog, a spherical type mechanical check device (SMCD) that is form of spherical differing from the existing form of flat or cube has been designed and tested its practicability to replace the part in mechanical check with digital image from QA operation. Materials and Methods: If the distance maintains constant between source(target) and image detector with constant distance to the center of spherical type mechanical check device(SMCD), the size will be shown as a constant image at all times regardless of its direction exposed. For the test, two accurate hemispheres are made and put together which results in a sphere of the equilateral circle. Results: It enables a variety of implementation of the existing mechanical check using digital image as follows: congruity level of radiation field and light field, size accuracy of radiation field and collimation field, gantry rotation isocenter check, collimation rotation isocenter check, room laser accuracy check, collimation rotation angle check, couch rotation angle check, and more. Conclusion: It has proved its practicability in checking isocenter congruity level as real time at the time of simultaneous rotation between gantry and couch that is applied to the non-coplanar field, which had been hard to apply as a device formed of existing flat or cube.

  • PDF

Feasibility Study of Robotics-based Patient Immobilization Device for Real-time Motion Compensation

  • Chung, Hyekyun;Cho, Seungryong;Cho, Byungchul
    • Progress in Medical Physics
    • /
    • v.27 no.3
    • /
    • pp.117-124
    • /
    • 2016
  • Intrafractional motion of patients, such as respiratory motion during radiation treatment, is an important issue in image-guided radiotherapy. The accuracy of the radiation treatment decreases as the motion range increases. We developed a control system for a robotic patient immobilization system that enables to reduce the range of tumor motion by compensating the tumor motion. Fusion technology, combining robotics and mechatronics, was developed and applied in this study. First, a small-sized prototype was established for use with an industrial miniature robot. The patient immobilization system consisted of an optical tracking system, a robotic couch, a robot controller, and a control program for managing the system components. A multi speed and position control mechanism with three degrees of freedom was designed. The parameters for operating the control system, such as the coordinate transformation parameters and calibration parameters, were measured and evaluated for a prototype device. After developing the control system using the prototype device, a feasibility test on a full-scale patient immobilization system was performed, using a large industrial robot and couch. The performances of both the prototype device and the realistic device were evaluated using a respiratory motion phantom, for several patterns of respiratory motion. For all patterns of motion, the root mean squared error of the corresponding detected motion trajectories were reduced by more than 40%. The proposed system improves the accuracy of the radiation dose delivered to the target and reduces the unwanted irradiation of normal tissue.

Evaluation of dose delivery accuracy due to variation in pitch and roll (세기변조방사선치료에서 Pitch와 Roll 변화에 따른 선량전달 정확성 평가)

  • Jeong, Chang Young;Bae, Sun Myung;Lee, Dong Hyung;Min, Soon Ki;Kang, Tae Young;Baek, Geum Mun
    • The Journal of Korean Society for Radiation Therapy
    • /
    • v.26 no.2
    • /
    • pp.239-245
    • /
    • 2014
  • Purpose : The purpose of this study is to verify the accuracy of dose delivery according to the pitch and roll rotational setup error with 6D robotic couch in Intensity Modulated Radiation Therapy (IMRT) for pelvic region in patients. Materials and Methods : Trilogy(Varian, USA) and 6D robotic couch(ProturaTM 1.4, CIVCO, USA) were used to measure and analyze the rotational setup error of 14 patients (157 setup cases) for pelvic region. The total 157 Images(CBCT 78, Radiography 79) were used to calculate the mean value and the incidence of pitch and roll rotational setup error with Microsoft Office Excel 2007. The measured data (3 mm, 3%) at the reference angle ($0^{\circ}$) without couch rotation of pitch and roll direction was compared to the others at different pitch and roll angles ($1^{\circ}$, $1.5^{\circ}$, $2^{\circ}$, $2.5^{\circ}$) to verify the accuracy of dose delivery by using 2D array ionization chamber (I'mRT Matrixx, IBA Dosimetry, Germany) and MultiCube Phantom(IBA Dosimetry, Germany). Result from the data, gamma index was evaluated. Results : The mean values of pitch and roll rotational setup error were $0.9^{\circ}{\pm}0.7$, $0.5^{\circ}{\pm}0.6$. The maximum values of them were $2.8^{\circ}$, $2.0^{\circ}$. All of the minimum values were zero. The mean values of gamma pass rate at four different pitch angles ($1^{\circ}$, $1.5^{\circ}$, $2^{\circ}$, $2.5^{\circ}$) were 97.75%, 96.65%, 94.38% and 90.91%. The mean values of gamma pass rate at four different roll angles ($1^{\circ}$, $1.5^{\circ}$, $2^{\circ}$, $2.5^{\circ}$) were 93.68%, 93.05%, 87.77% and 84.96%. when the same angles ($1^{\circ}$, $1.5^{\circ}$, $2^{\circ}$) of pitch and roll were applied simultaneously, The mean values of each angle were 94.90%, 92.37% and 87.88%, respectively. Conclusion : As a result of this study, it was able to recognize that the accuracy of dose delivered is lowered gradually as pitch and roll increases. In order to increase the accuracy of delivered dose, therefore, it is recommended to perform IGRT or correct patient's position in the pitch and roll direction, to improve the quality of treatment.

Real-time Body Surface Motion Tracking using the Couch Based Computer-controlled Motion Phantom (CBMP) and Ultrasonic Sensor: A Feasibility Study (CBMP (Couch Based Computer-Controlled Motion Phantom)와 초음파센서에 기반한 실시간 체표면 추적 시스템 개발: 타당성 연구)

  • Lee, Suk;Yang, Dae-Sik;Park, Young-Je;Shin, Dong-Ho;Huh, Hyun-Do;Lee, Sang-Hoon;Cho, Sam-Ju;Lim, Sang-Wook;Jang, Ji-Sun;Cho, Kwang-Hwan;Shin, Hun-Joo;Kim, Chul-Yong
    • Progress in Medical Physics
    • /
    • v.18 no.1
    • /
    • pp.27-34
    • /
    • 2007
  • Respiration sating radiotherapy technique developed In consideration of the movement of body surface and Internal organs during respiration, is categorized into the method of analyzing the respiratory volume for data processing and that of keeping track of fiducial landmark or dermatologic markers based on radiography. However, since these methods require high-priced equipments for treatment and are used for the specific radiotherapy. Therefore, we should develop new essential method whilst ruling out the possible problems. This study alms to obtain body surface motion by using the couch based computer-controlled motion phantom (CBMP) and US sensor, and to develop respiration gating techniques that can adjust patients' beds by using opposite values of the data obtained. The CBMP made to measure body surface motion is composed of a BS II microprocessor, sensor, host computer and stopping motor etc. And the program to control and operate It was developed. After the CBMP was adjusted by entering random movement data, and the phantom movements were acquired using the sensors, the two data were compared and analyzed. And then, after the movements by respiration were acquired by using a rabbit, the real-time respiration gating techniques were drawn by operating the phantom with the opposite values of the data. The result of analysing the acquisition-correction delay time for the data value shows that the data value coincided within 1% and that the acquistition-correction delay time was obtained real-time $(2.34{\times}10^{-4}sec)$. And the movement was the maximum movement was 6 mm In Z direction, In which the respiratory cycle was 2.9 seconds. This study successfully confirms the clinical application possibility of respiration gating techniques by using a CBWP and sensor.

  • PDF

Modified Five Field Technique for Primary and Postop Breast Cancer Irradiation (유방암에서의 근치적 또는 수술후 방사선 치료방법 : 5문 조사법)

  • Choi, Eun-Kyung;Chang, Hye-Sook;Yi, Byong-Yong
    • Radiation Oncology Journal
    • /
    • v.9 no.1
    • /
    • pp.165-170
    • /
    • 1991
  • In breast cancer, the treatment volume presents a relatively complex three dimensional structure. Effective radiation therapy requires the delivery of adequate dose to a large target volume using complex beam arrangements. The technique proposed here is our department's method using asymmetric jaw with appropriate couch, collimator and gantry rotation. This technique has the following advantages: 1) all treatments are given in a single clinical set up 2) it does not require half beam blocks 3) it produces exact geomatric match 4) it is very convenient and easy to use 5) it has daily reproducibility.

  • PDF

Evaluation of the Usefulness of MapPHAN for the Verification of Volumetric Modulated Arc Therapy Planning (용적세기조절회전치료 치료계획 확인에 사용되는 MapPHAN의 유용성 평가)

  • Woo, Heon;Park, Jang Pil;Min, Jae Soon;Lee, Jae Hee;Yoo, Suk Hyun
    • The Journal of Korean Society for Radiation Therapy
    • /
    • v.25 no.2
    • /
    • pp.115-121
    • /
    • 2013
  • Purpose: Latest linear accelerator and the introduction of new measurement equipment to the agency that the introduction of this equipment in the future, by analyzing the process of confirming the usefulness of the preparation process for applying it in the clinical causes some problems, should be helpful. Materials and Methods: All measurements TrueBEAM STX (Varian, USA) was used, and a file specific to each energy, irradiation conditions, the dose distribution was calculated using a computerized treatment planning equipment (Eclipse ver 10.0.39, Varian, USA). Measuring performance and cause errors in MapCHECK 2 were analyzed and measured against. In order to verify the performance of the MapCHECK 2, 6X, 6X-FFF, 10X, 10X-FFF, 15X field size $10{\times}10$ cm, gantry $0^{\circ}$, $180^{\circ}$ direction was measured by the energy. IGRT couch of the CT values affect the measurements in order to confirm, CT number values : -800 (Carbon) & -950 (COUCH in the air), -100 & 6X-950 in the state for FFF, 15X of the energy field sizes $10{\times}10$, gantry $180^{\circ}$, $135^{\circ}$, $275^{\circ}$ directionwas measured at, MapPHAN allocated to confirm the value of HU were compared, using the treatment planning computer for, Measurement error problem by the sharp edges MapPHAN Learn gantry direction MapPHAN of dependence was measured in three ways. GANTRY $90^{\circ}$, $270^{\circ}$ in the direction of the vertically erected settings 6X-FFF, 15X respectively, and Setting the state established as a horizontal field sizes $10{\times}10$, $90^{\circ}$, $45^{\circ}$, $315^{\circ}$, $270^{\circ}$ of in the direction of the energy-6X-FFF, 15X, respectively, were measured. Without intensity modulated beam of the third open arc were investigated. Results: Of basic performance MapCHECK confirm the attenuation measured by Couch, measured from the measured HU values that are assigned to the MAP-PHAN, check for calculation accuracy for the angled edge of the MapPHAN all come in a range of valid measurement errors do not affect the could see. three ways for the Gantry direction dependence, the first of the meter built into the value of the Gantry $270^{\circ}$ (relative $0^{\circ}$), $90^{\circ}$ (relative $180^{\circ}$), 6X-FFF, 15X from each -1.51, 0.83% and -0.63, -0.22% was not affected by the AP/PA direction represented. Setting the meter horizontally Gantry $90^{\circ}$, $270^{\circ}$ from the couch, Energy 6X-FFF 4.37, 2.84%, 15X, -9.63, -13.32% the difference. By-side direction measurements MapPHAN in value is not within the valid range can not, because that could be confirmed as gamma pass rate 3% of the value is greater than the value shown. You can check the Open Arc 6X-FFF, 15X energy, field size $10{\times}10$ cm $360^{\circ}$ rotation of the dose distribution in the state to look at nearly 90% pass rate to emerge. Conclusion: Based on the above results, the MapPHAN gantry direction dependence by side in the direction of the beam relative dose distribution suitable for measuring the gamma value, but accurate measurement of the absolute dose can not be considered is. this paper, a more accurate treatment plan in order to confirm, Reduce the tolerance for VMAT, such as lateral rotation investigation in order to measure accurate absolute isodose using a combination of IMF (Isocentric Mounting Fixture) MapCHEK 2, will be able to minimize the impact due to the angular dependence.

  • PDF

Dosimetric Characteristics of Multileaf Collimator-based Intensity-modulated Arc Therapy for Stereotactic Radiosurgery (방사선수술 시 다엽 콜리메이터를 기초로 한 IMAT의 선량분포)

  • Yun, Sang-Mo;Kim, Sung-Kyu
    • Progress in Medical Physics
    • /
    • v.18 no.2
    • /
    • pp.93-97
    • /
    • 2007
  • This study was designed to evaluate radiosurgery technique using multiple noncoplanar arc therapy with intensity modulated fine MLC shaped photon beam. The stereotactic radiosurgery was performed with 6-MV X-ray beams from a Clinac 21EX LINAC (Varian, Palo Alto, CA, USA) with a MLC-120, which features a full $40{\times}40cm$ field and is the first MLC for general use that offers 0.5 cm resolution for high precision treatment of small and irregular fields. We used a single isocenter and five gantry-couch combinations with a set of intensity modulated arc therapy. We investigated dosimetric characteristics of 2 cm sized spherical target volume with film (X-OMAT V2 film, Kodak Inc, Rochester NY, USA) dosimetry within $25{\times}25cm$ acrylic phantom. A simulated single isocentric treatment using inversely Planned 3D radiotherapy planning system demonstrated the ability to conform the dose distribution to an spherical target volume. The 80% dose level was adequate to encompass the target volume in frontal, sagittal, and transverse planes, and the region between the 40% and 80% isodose lines was $4.0{\sim}4.5mm$ and comparable to the dose distribution of the Boston Arcs. We expect that our radiosurgery technique could be a treatment option for irregular-shaped large intracranial target.