• 제목/요약/키워드: Tear Sampling

검색결과 1건 처리시간 0.013초

눈물의 생체 분자 추출을 위한 미세유관이 탑재된 유연한 콘택트 렌즈 (Micro-channel Embedded Soft Contact Lens for Tear Biomolecule Sampling)

  • 김옥철;박형달;송용원;김진석
    • 대한의용생체공학회:의공학회지
    • /
    • 제37권5호
    • /
    • pp.178-185
    • /
    • 2016
  • Tear is a promising biological fluid for non-invasive health monitoring. It has been studied in the past to be a possible candidate for the diagnosis of certain systemic diseases, such as breast cancer, multiple sclerosis, prostate cancer, and diabetes. However, currently existing methods for collecting and extracting tear from the human eye causes inconsistencies in the biomolecule concentrations of the tear sample due to the irritating nature of the process. In response, we designed and fabricated a microfluidic system embedded soft contact-lens for the purpose of tear sampling. The lens was then tested with artificial tear for its tear sampling capability, and found to be able to find concentration equilibrium within 50 minutes. Additionally, simulation was carried out to further optimize the design so that tear sampling rate matched the natural tear turn-over rate of 1 microliter per minute.