• Title/Summary/Keyword: Silver Halide Optical Fiber

Search Result 4, Processing Time 0.021 seconds

Development of Noncontact Temperature Sensor Using Silver Halide Optical Fiber for Medical Usages (Silver Halide 광섬유를 이용한 의료용 비접촉식 온도 센서 개발)

  • Yoo, Wook-Jae;Cho, Dong-Hyun;Jang, Kyoung-Won;Chung, Soon-Cheol;Tack, Gye-Rae;Lee, Bong-Soo
    • Journal of Biomedical Engineering Research
    • /
    • v.27 no.6
    • /
    • pp.337-342
    • /
    • 2006
  • We have developed a noncontact temperature sensor using a silver halide optical fiber. The infrared collimator and focus head are connected both ends of a silver halide optical fiber with SMA connectors and used to collimate radiations of a heat source and to focus them to infrared sensors such as a pyroelectric sensor and a thermopile sensor, respectively. The relation ships between the temperatures of a heat source and the output signals of the infrared sensors are determined to measure the surface temperature of a heat source. The measurable temperature range is from 25 to $60^{\circ}C$. It is expected that a noncontact temperature sensor using a silver halide optical fiber can be developed for medical usages such as temperature monitoring during hyperthermia, cryosurgery, laser surgery and diagnostic procedure based on the results of this study.

Development of a Fiber-optic Noncontact Temperature Sensor for Measuring the Temperature of Cooled Secondary Water in a Nuclear Power Plant (냉각된 원전 2차계통수의 온도측정을 위한 비접촉식 광섬유 온도센서의 개발)

  • Yoo, Wook-Jae;Lee, Bong-Soo;Park, Byung-Gi;Cho, Young-Ho
    • Journal of the Korea Academia-Industrial cooperation Society
    • /
    • v.11 no.5
    • /
    • pp.1730-1734
    • /
    • 2010
  • Generally, a pH value of secondary water in a nuclear power plant should be estimated after sampling and cooling down. In this process, the measurement of temperature is very important because a pH value is varied according to the temperature of secondary water. In this study, a noncontact fiber-optic temperature sensor using a silver halide optical fiber is fabricated to measure the temperature of cooled secondary water. And we have measured an infrared radiation, which is transferred by a silver halide optical fiber from a heat source, using a thermopile sensor. The relationships between the temperature of a heat source and the output voltage of the fiber-optic temperature sensor according to the change of distance and angle are determined. The measurable temperature range of the fiber-optic temperature sensor is from 25 to $60^{\circ}C$. Based on the results of this study, a noncontact temperature sensor using a silver halide optical fiber can be developed for the temperature measurement of the pH sample in the secondary water system.

Measurements of temperature distribution using an infrared optical fiber during radiofrequency ablation (적외선 투과 광섬유를 이용한 고주파 열치료 과정에서의 온도분포 측정)

  • Yoo, Wook-Jae;Seo, Jeong-Ki;Cho, Dong-Hyun;Jang, Kyoung-Won;Shin, Sang-Hun;Lee, Bong-Soo;Tack, Gye-Rae;Park, Byung-Gi;Moon, Joo-Hyun
    • Journal of Sensor Science and Technology
    • /
    • v.17 no.6
    • /
    • pp.397-405
    • /
    • 2008
  • In this study, we have measured temperature distribution using infrared optical fibers during radiofrequency ablation (RFA). Infrared radiations generated from the water around inserted electrode are transferred by silver halide optical fibers and are measured by a thermopile sensor. Also, the output voltages of a thermopile sensor are compared with those of the thermocouple recorder. It is expected that a noncontact temperature sensor using an infrared optical fiber can be developed for the temperature monitoring during RFA treatments based on the results of this study.