• Title/Summary/Keyword: Resonant inverter type X-ray generator

Search Result 6, Processing Time 0.02 seconds

Design of High Frequency Resonant Inverter Type X-Ray Generator (고주파 공진형 인버터식 X선 발생장치의 설계)

  • Lee, Seong-Gil;Park, Su-Gang;Baek, Hyeong-Rae;Jeong, Su-Bok
    • The Transactions of the Korean Institute of Electrical Engineers B
    • /
    • v.51 no.1
    • /
    • pp.34-39
    • /
    • 2002
  • Most X-ray generator had been used do rectifier type transformer high tension generator which is supplied in a clinical diagnosis. But it is difficult to miniaturize and become light weight. Also, because the ripple rate of tube voltage is high, X-ray generating efficiency is very low. Therefore, it is supplied gradually from abroad being developed high tension generator for inverter type X-ray generator which use semi-conductor switching element for electric power that have high speed switching ability to solved these problem. But, semi-conductor element of big capacity are used by X-ray tube's big consumption power and diffusion is difficult in the small size hospital because production cost is ascending by doing digital control through DSP and product price becomes expensive. Therefore, in this paper, design and manufactured CR type voltage divider for feedback control of tube voltage of high frequency resonance type inverter and high tension transformer for high frequency to apply economical diffusion type X-ray generator which have wide output voltage and load extent. It is Proved do X-ray generator and stability of X-ray tube's output characteristics through an experiment.

A Study on the Output Characteristics Comparison of High Frequency Resonant Inverter Type X-ray Generators in Short Exposure Time (고주파 공진형 인버터식 X-선 장치의 단시간 출력특성 비교 연구)

  • 정수복;이성길;임홍우;백형래
    • The Transactions of the Korean Institute of Power Electronics
    • /
    • v.4 no.1
    • /
    • pp.66-72
    • /
    • 1999
  • This paper deals with the output characteristics of resonant PWM inverter type X-ray generators connected to different DC power units i.e. a single phase full bridge rectifier, a three phase full bridge rectifier and a power storage unit(PSU). The quality of X-ray beam depends on the pulsating waveforms of DC voltage supplied to the X-ray tube. In a X-ray generator, the waveform of DC output voltage can be affected from affected from harmonic distortion of DC input power. When a tube voltage waveform is distorted, the property of X-ray beam such as reproducibility, linearity and dose can be reduced. Therefore, we compared DC output waveforms and dose with three type of DC power units and show the experimental results in this paper.

Design of High Frequency Resonant High-tension Transformer in Inverter Type X-Ray Generator (인버터식 X선발생장치용 고주파 공진형 고압변압기 등의 설계)

  • Lee, Seong-Kil;Choi, Sung-Kwan
    • Journal of radiological science and technology
    • /
    • v.24 no.2
    • /
    • pp.5-11
    • /
    • 2001
  • Most of X-ray generator had used rectifier type transformer with high tension generator which is supplied in a clinical diagnosis. Because the ripple rate of tube voltage is high, X-ray generating efficiency is very low. In these days, high tension generator for inverter type X-ray generator is being supplied from a broad which uses semi-conductor switching element for the electric power that have a high speed switching ability to solve these problem. But, semi-conductor element with large capacity is used with X-ray tube's large consumption power and diffusion is difficult in the small size hospital because production cost is going up by doing digital control through DSP. Therefore, this paper designed and manufactured CR type voltage divider for feedback control of tube voltage with high frequency resonance type inverter and for high tension transformer with high frequency. It was to make economical diffusion type X-ray generator which has wide output voltage and load extent. It was preyed that the X-ray generator had the stability of X-ray tube's output characteristics.

  • PDF

A Study on the Output Characteristics of High-Frequency Resonant Inverter Type for X-ray Generators in Short Exposure Time. (고주파 공진형 인버터 X선 장치의 단시간 출력특성에 관한 연구)

  • 이성길;임홍우;이상일;조금배;백형래
    • Proceedings of the KIPE Conference
    • /
    • 1998.07a
    • /
    • pp.129-133
    • /
    • 1998
  • This paper deals with a characteristic of output for high frequency resonant inverter type x-ray generator in short exposure time. In the same tube-voltage, the waveform is stable in long exposure time (200msec) but more decreasing exposure time (50msec), the waveform is distorted. The ripple factor of tube-voltage waveform distorted more and more increase tube-voltage and in this case output is also unstable. High-frequency resonant inverter type X-ray generator using PSU source which introduced in resent is stable tube-voltage waveform and low ripple factor.

  • PDF

THE CHARACTERISTICS OF OUTPUT FOR INVERTER TYPE X-RAY GENERATOR

  • Lim, Hong-Woo;Han, Euam-Yong;Baek, Hyung-Lae;Lee, Seong-Kil
    • Proceedings of the KIPE Conference
    • /
    • 1998.10a
    • /
    • pp.431-435
    • /
    • 1998
  • This paper deals with the output characteristics of resonant PWM inverter type X-ray generators connected to different DC power units i.e. a single phase full bridge rectifier, a three phase full bridge rectifier and a power storage unit(PSU). The quality of X-ray beam depend on the pulsation waveforms of DC voltage supplied to the X-ray tulbe. In a X-ray generator, the waveforme of DC output voltage can be affected from hramonic distortion of DC input power. When a tube voltage waveform is distorded, the property of X-ray beam such as reproducibility, direcibility and doesage can be reduced. Therefore, we compared DC output waveforms and doesages with three thpe of DC power units and show the experimental results in this paper

  • PDF

A 37.5kV A Class Power Supply for Diagnostic X -ray by High-Frequency Resonant Inverter Type (37.5kVA급 고주파 공진형 인버터식 진단 X선 전원장치)

  • 김학성;유동욱;임근희;완충연
    • Journal of the Korean Institute of Telematics and Electronics B
    • /
    • v.33B no.12
    • /
    • pp.113-121
    • /
    • 1996
  • This paper presents a 37.5kVA (125kVp, 300mA) high frequency inverter type power supply for diagonostic X-ray generator with two HFHV (high-frequency high-voltage) isolating transformers. The output regulations, X-ray tuve voltage and tube current, are obtained by two high frequency inverters with the each other feedback PI controller and operating time switch for selecting shooting time and synchronizing two inverters. The operating modes and design consideration of the proposed power supply are given. Issues in the design of two HFHV isolating transformersfors high volage genrating and filament heating are discussed. Experimental results are presented to verify the performance of the designed power supply for varying load conditions. The proposed apparatus has several advnatages, e.g., the fast rising time of tube voltae, accuracy and reduced component size etc.

  • PDF