• Title/Summary/Keyword: Record Management Personnel

Search Result 41, Processing Time 0.018 seconds

Analyze dosimetry with and without shielding when amplifying scattered rays (산란선 증폭시 차폐체 유무에 따른 선량 분석)

  • Chang Ho Cho;Jeong Lae Kim
    • The Journal of the Convergence on Culture Technology
    • /
    • v.10 no.3
    • /
    • pp.819-825
    • /
    • 2024
  • The reason for recording dose data when using a diagnostic radiation source is to record and manage the dose to healthcare personnel and patients. The purpose of this study was to verify the difference in radiation dose when using diagnostic radiation generating devices and to inform users' awareness of dose reduction through measurement and analysis of dose in situations with and without shielding. The dose analysis of each equipment for two Korean C-arms and two German C-arms showed that the Korean FPD type C-arm had the highest dose value, followed by the German I.I type C-arm, German FPD type C-arm, Korean, and I.I type C-arm. The results of the dose analysis with and without shielding showed that the dose to the human phantom in a normal atmosphere increased by about 2 times due to scattered radiation, but the dose to the human phantom was reduced by about 5 times by wearing a shield (0.5mm/lead apron). More important than the management of radiation dose is the study of how to reduce exposure when using radiation, and since the radiation dose output from different equipment is different, it is necessary to provide dose information with and without shielding.