Comparative Evaluation of First Urine and Intermediate Urine Samples Collected using a Patented Urine Cup (특허 받은 소변컵을 이용하여 채집한 첫 소변과 중간소변 시료의 비교 평가)

  • Kim, SeungChul;Kim, HoSung;Kim, ChangUook;Pyo, SangShin
    • The Journal of the Korea Contents Association
    • /
    • v.22 no.6
    • /
    • pp.619-628
    • /
    • 2022
  • In general, as a method to confirm a urinary tract infection (UTI) in a medical institutions, urine culture including a urinalysis and an antimicrobial susceptibility test is performed. It is important to disinfect the area around the urethra and perineum before collecting urine samples, and it is important to collect it intermediate urine, not the first-void urine. We invented a patent urine cup (Patent No. 10-1732843) that can automatically and easily separate first-void urine and midstream urine and using this, the patent cup and the general cup were compared and evaluated using this. Nitrite (P<0.001), WBC (P=0.005), Bacterial colony count (P=0.001), colony positivity rate (P=0.004) in first-void urine (N=24), midstream urine (N=24) separated by patent cup to obtain a significantly higher value. This can be seen from the fact that the first-void urine and midstream urine separated using the patent cup were well separated. Also, the number of Bacterial colonies was statistically significantly higher in the midstream urine isolated using a patent cup (N=24) than in the midstream urine collected using a general cup (N=24) (average 7.9 vs. 4.0 on average, P= 0.002). Which means that the midstream urine separated using the patent cup is more sensitive to the UTI test than the midstream urine collected using a general cup.