• Title/Summary/Keyword: PC camera$KMnO_4$ consumption

Search Result 1, Processing Time 0.013 seconds

Automatic Titration for KMnO4 Consumption Test of Tap Water Using Personal Computer Camera (PC 카메라를 이용한 수돗물의 과망간산칼륨소비량 적정 자동화)

  • Lee, Hyeong-Choon
    • Journal of Environmental Health Sciences
    • /
    • v.34 no.1
    • /
    • pp.95-100
    • /
    • 2008
  • An automatic titration system using a PC-camera with a color filter on its lens was used in the $KMnO_4$ consumption test of tap water and distilled water in relation to blank tests. The very faint pink color of titration end point could be effectively detected by using a yellow cellophane paper as a color filter. The average hue value (Havg) of 192 pixels in the image of the sample solution being titrated was computed and followed up at regular time intervals during titration in order to detect the titration end point. The Havg decrease of 2 degrees from the average of first 10 Havgs was regarded as reaching the end point. The volume of 0.01N $KMnO_4$ consumed by a tap water sample was $0.728{\pm}0.022ml$ in manual titration and $0.735{\pm}0.013ml$ in automatic titration (p=0.580). The volume of 0.01N $KMnO_4$ consumed by a distilled water sample was $0.383{\pm}0.015ml$ in manual titration and $0.367{\pm}0.015ml$ in automatic titration (p=0.252). The high p-values for t-test suggested that there were good agreements between manual and automatic titration data and the automatic method proposed in this article was considered to effectively replace the manual titration.