• Title/Summary/Keyword: Oxylator EM-100

Search Result 2, Processing Time 0.014 seconds

Comparison of the Ventilatory Volume and Airway Pressures using Oxylator EM-100, MicroVenT CSI-3000, OXY-LIFE II -A manikin study - (산소소생기 Oxylator EM-100, MicroVenT CSI-3000, OXY-LIFE II의 환기량과 기도내압 비교 - 마네킨 연구 -)

  • Shim, Gyu-Sik;Kim, Eun-Mee;Roh, Sang-Gyun
    • Fire Science and Engineering
    • /
    • v.31 no.2
    • /
    • pp.113-118
    • /
    • 2017
  • The purpose of this study was to compare the ventilatory volume and airway pressure using Oxylator EM-100, MicroVenT CSI-3000, OXY-LIFE II. The data were obtained from February 13 in 2017 andanalyzed using the SPSS WIN 18.0 program. The results of theventilatory volume showed the Oxylator EM-100 551.44 ml (${\pm}18.70$), MicroVenT CSI-3000 527.26 ml (${\pm}17.98$), and OXY-LIFE II 369.46 ml (${\pm}12.30$). The airway pressure showed the Oxylator EM-100 $11.89cmH_2O$ (${\pm}.41$), MicroVenT CSI-3000 $11.66cmH_2O$ (${\pm}.34$), and OXY-LIFE II $8.02cmH_2O$ (${\pm}.25$). This study will provide the basic data for an appropriate ventilation method by an oxygen supply device including Oxylator EM-100, MicroVenT CSI-3000, and OXY-LIFE II.

Comparison of Ventilatory Volume and Airway Pressures Using Oxylator EM-100 (옥시레이터 EM-100을 활용한 환기량과 기도내압 비교)

  • Shin, So-Yeon;Roh, Sang-Gyun
    • Fire Science and Engineering
    • /
    • v.29 no.5
    • /
    • pp.104-109
    • /
    • 2015
  • The purpose of this study was to compare the ventilatory volume and airway pressure of a facial mask, endotracheal intubation, King tube, and I-gel devices with an Oxylator EM-100 using a RespiTrainer. The data were obtained from July 20 to 21, 2015. Data were analyzed using SPSS WIN 18.0 software. The ventilatory volume for endotracheal intubation was 537 ml (95% CI 530~545 ml), that for the King tube was 502 ml (95% CI 499~506 ml), that for the I-gel was 88 ml (95% CI 485~491 ml), and that for the facial mask was 499 ml (95% CI 496~503 ml). The airway pressure for endotracheal intubation was $11.34cmH_2O$ (95% CI $11.21{\sim}11.41cmH_2O$), that for the King tube was $10.67cmH_2O$ (95% CI $10.60{\sim}10.75cmH_2O$), that for the I-gel was $10.42cmH_2O$ (95% CI $10.35{\sim}10.67cmH_2O$), and that for the facial mask was $10.61cmH_2O$ (95% CI $10.55{\sim}10.68cmH_2O$). As a result, we were able to identify the appropriate ventilatory volume using the Oxylator EM-100.