Evaluation of Dosimetric Characteristics of Reproducibility, Linearity and Dose Dependence of Optically Stimulated Luminescence Dosimeters in Co-60 Gamma-rays (Co-60 감마선을 이용한 광자극발광선량계의 재현성, 선형성, 선량의존성에 대한 특성평가)

  • Han, Su Chul;Choi, Sang Hyoun;Park, Seungwoo;Kim, Chul Hang;Jung, Haijo;Kim, Mi-Sook;Yoo, Hyung Jun;Kim, Chan Hyeong;Ji, Young Hoon;Yi, Chul Young;Kim, Kum Bae
    • Progress in Medical Physics
    • /
    • v.25 no.1
    • /
    • pp.31-36
    • /
    • 2014
  • We aimed to evaluate the dosimetric characteristics of reproducibility, linearity and dose dependence of optical stimulated luminance dosimeter (OSLD) in the Co-60 Gamma-rays and to analyze with a precedent study in field of the diagnostic radiography and radiotherapy. The reproducibility was 0.76% of the coefficient of variation, the homogeneity was within 1.5% of the coefficient of variation and OSLD had supra-linear response more than 3 Gy. So the correlation between dose and count was fitted by quadratic function. The count depletion by repeated reading was 0.04% per reading regardless of the irradiated dose. And the half time of decay curve according to the irradiated dose was 0.68 min. with 1 Gy, 1.04 min. with 5 Gy, and 1.10 min. with 10 Gy, respectively. In case of annealing for 30 min, the removal rate was 88% with 1 Gy, 90% with 5 Gy, and 92% with 10 Gy, respectively and 99% in case of annealing time for 4 hour. It is feasible to use OSLDs for dose evaluation in Co-60 Gamma-rays when considering the uncertainty on the procedure according to the irradiated dose.