• Title/Summary/Keyword: One-piece abutment

Search Result 16, Processing Time 0.017 seconds

Study of screw loosening in cementation type implant abutment

  • Hwang, Bo-Yeon;Kim, Yung-Soo;Kim, Chang-Whe
    • The Journal of Korean Academy of Prosthodontics
    • /
    • v.38 no.6
    • /
    • pp.765-781
    • /
    • 2000
  • The purpose of this study was to compare the screw loosening characteristics of three avail-able cementation type abutments: one-piece cementation type abutment; two-piece cementation type abutment using titanium abutment screw; two-piece cementation type abutment using gold abutment screw. Two implant supported three-unit superstructures were fabricated using a pair of 3 kinds of abutments for each experimental model. Cyclic loading was applied on the specimen, and made to stop when the superstructure showed movement over threshold range. The loaded cycle was counted until the machine stopped. Frequency analysis was done to measure the change of natural frequency before and after the application of cyclic load and to find the effect of screw loosening on the change of natural frequency. The specimen assembly was modeled to perform the finite element analysis to see the distribution of the stress induced by the application of preload over the screw joint and to compare the pattern of the distribution of stress induced by the external force with the change of the preload condition. The following results were obtained: 1. The failure loading cycle of two-piece cementation type abutment using gold screw was significantly greater than those of the other groups. 2. One-piece cementation type abutment applied to multi-unit restoration case did not show greater resistance to screw loosening compared to two-piece cementation type abutments. 3. Frequency analysis showed decrease in natural frequency when screw loosening occured.

  • PDF

Considerations in implant crestal module to preserve peri-implant tissue (임플란트 주위 조직 보존을 위한 임플란트 경부의 디자인에 관한 고찰)

  • Kim, Hong-Jun;Kim, Jee-Hwan;Kim, Sung-Tae;Lee, Jae-Hoon;Park, Young-Bum
    • The Journal of Korean Academy of Prosthodontics
    • /
    • v.49 no.4
    • /
    • pp.346-353
    • /
    • 2011
  • Purpose: The peri-implant soft tissue is remodeled by the initial marginal bone resorption affecting the prognosis and esthetic result of treatment. Thus various designs on implant neck design are studied to preserve peri-implant bone. The purpose of this study is to review on the causes of initial marginal bone resorption, the configuration of peri-implant soft tissue, and the implant crestal module favorable in preserving peri-implant tissue. Materials and methods: The studies on the causes of initial marginal bone resorption and the implant crestal modules are researched and reviewed using Pubmed database. The implant crestal modules including one piece and two-piece implant, internal and external hex abutment, taper and butt joint connection, scalloped design abutment, and platform switching concept are reviewed. Results: A number of clinical and experimental studies preferred one piece implant to two-piece in preserving initial peri-implant tissue. For two piece implants, internal hex abutment and taper joint connection appear more favorable than external hex abutment and butt joint connection relatively. Controversial issues still exist on scalloped design requiring more studies on it. Although the rationale is not certain, the concept of platform switching seems favorable in preserving initial peri-implant tissue based on clinical and experimental studies. Conclusion: Each implant crestal module contains its own advantages and disadvantages with various controversial issues. In the aspect of preservation of initial peri-implant tissue, however, one-piece implant seems beneficial. In cases when two-piece implant is more appropriate due to prosthodontic concerns or any other problems, the application of platform switching concept, internal connection abutment, and taper joint connection may be favorable for the preservation of peri-implant tissues.

Mandibular 4 incisors implant restoration (하악 4전치 상실 시 임플란트 수복)

  • Park, Jong hyun
    • Journal of the Korean Academy of Esthetic Dentistry
    • /
    • v.31 no.2
    • /
    • pp.56-63
    • /
    • 2022
  • The mandibular 4 anterior incisor restoration has small teeth and a small space, so the choice of abutment is always a concern. The mandibular 4 anterior incisor extraction is caused by periodontal disease, interdental embrasure greatly open is advantageous of oral hygiene management. Try to make it small diameter with custom abutment, it is limited to the space for the screw. Rather than setting the post site of one body implant to a horizontal cross-section, it is advantageous to set it to a longitudinal cross-section, for interdental embrasure formation. When using an internal bone level implant, rather than using a two-piece abutment, using a one-piece abutment can more effectively secure space for interdental embrasure.

EFFECT OF A COUNTER-TORQUE DEVICE AND THE INTERNAL HEXAGON OF ABUTMENT ON THE TIGHTENING TORQUE TRANSMITTED TO THE IMPLANT (회전방지장치와 지대주의 내육각구조가 임플란트로 전달되는 조임 회전력에 미치는 영향)

  • Lee Sang-Min;Jeon Young-Chan;Jeong Chang-Mo
    • The Journal of Korean Academy of Prosthodontics
    • /
    • v.41 no.2
    • /
    • pp.223-231
    • /
    • 2003
  • Statement of problem : Little is known about the effect of a counter-torque device and the internal hexagon of abutment on the tightening torque transmitted to the implant. Purpose : The purpose of this study was to examine the effect of a counter-torque device and the internal hexagon of abutment on the tightening torque transmitted to the implant. Material and Methods : In this study, three types of abutment were used, (1) two-piece conical abutment with hexagon, (2) two-piece conical abutment without hexagon, and (3) one-piece conical abutment without hexagon. The experimental groups were divided into five groups according to the type of abutment and the usage of a counter-torque device. Group I : two-piece conical abutment with internal hexagon was tightened without the use of a counter-torque device. Group II : two-piece conical abutment without internal hexagon was tightened without the use of a counter-torque device. Group III : one-piece conical abutment without internal hexagon was tightened without the use of a counter-torque device. Group IV : two-piece conical abutment with internal hexagon was tightened with the use of a counter-torque device Group V : two-piece conical abutment without internal hexagon was tightened with the use of a counter-torque device. Abutments were tightened 20Ncm torque with the use of manual torque wrench and then torque values were measured by torque-gauge. After the measurement of torque values, all groups were loosened with the use of manual torque wrench and then detorque values were measured by torque-gauge. Results : The results were as follows. 1. There were no differences in torque values among three types of abutment. 2. Regardless of the existence of the internal hexagon of abutment, a counter-torque device decreased the tightening torque transmitted to the implant about 92% 3. In group III showed the highest detorque value, however there were no differences among group I, II, IV and V. Conclusion : Within the limitations of this study, it was concluded that the internal hexagon of abutment has no effect on the tightening torque transmitted to the implant and the detorque value of abutment screw. The use of a counter-torque device is essential to prevent microfracture on the implant-bone interface but has no effect on preload.

FIT OF FIXTURE/ABUTMENT/SCREW INTERFACES OF INTERNAL CONNECTION IMPLANT SYSTEM

  • Kim, Jin-Sup;Kim, Hee-Jung;Chung, Chae-Heon;Baek, Dae-Hwa
    • The Journal of Korean Academy of Prosthodontics
    • /
    • v.43 no.3
    • /
    • pp.338-351
    • /
    • 2005
  • Statement of problem. Accurate fit between the implant components is important because the misfit of the implant components results in frequent screw loosening, irreversible screw fracture, plaque accumulation, poor soft tissue reaction, and destruction of osseointegration. Purpose. This study is to evaluate the machining accuracy and consistency of the implant fixture/ abutment/screw interfaces of the internal connection system by using a Stereoscopic Zoom microscope and FE-SEM(field emission scanning electron microscope) Materials and methods. The implant systems selected in this study were internal connection type implants from AVANA(Osstem^{\circledR}), Bioplant(Cowell-Medi^{\circledR}), Dio(DIO^{\circledR}), Neoplant(Neobiotech 􀋓), Implantium(Dentium􀋓)systems. Each group was acquired 2 fixtures at random. Two piece type abutment and one piece type abutment for use with each implant system were acquired. Screw were respectively used to hold a two piece type abutment to a implant fixture. The implant fixtures were perpendiculary mounted in acrylic resin block. Each two piece abutment was secured to the implant fixture by screw and one piece abutment also secured to the implant fixture. Abutment/fixture assembly were mounted in liquid unsaturated polyester. All samples were cross-sectioned with grinder-polisher unit. Finally all specimens were analysed the fit between implant fixture/abutment/screw interfaces Results and conclusions. 1. Implant fixture/abutment/screw connection interfaces of internal connection systems made in Korea were in good condition. 2. The results of the above study showed that materials and mechanical properties and quality of milling differed depending on their manufacturing companies.

Removal torque of a two-piece zirconia abutment with a novel titanium component in an internal connection implant system after dynamic cyclic loading (새로운 타이타늄 구성요소를 사용한 내부연결 임플란트용 지르코니아 지대주의 동적하중 후 나사 제거력)

  • Lee, Joo-Hee
    • The Journal of Korean Academy of Prosthodontics
    • /
    • v.55 no.2
    • /
    • pp.151-155
    • /
    • 2017
  • Purpose: The aim of this study was to evaluate screw removal torque of the two-piece zirconia abutment with the novel titanium component compared to the conventional one-piece titanium abutment in the internal connection implant before and after dynamic cyclic loading. Materials and methods: Two types of the abutment assemblies with internal connection were prepared and divided into the groups; titanium abutment-titanium abutment screw assemblies as control, and zirconia abutment-titanium socket-titanium abutment screw assemblies as experimental group. A total of 12 abutments and implants were used of six assemblies each group. Each assembly was tightened to 30 Ncm. A cyclic load of 300 N at an angle of 30 degrees in reference to the loading axis was applied until one million cycles or failure. The removal torque values (RTVs) of the abutment screws were measured with a digital torque gauge before and after cyclic loading. The RTVs of the pre load and post load were analyzed with t-test, and P-values < .05 were considered statistically significant. Results: The assemblies of both groups survived all after the dynamic cyclic loading test without screw loosening. The statistically significant differences were found between the mean RTVs before and after the cyclic loading in both groups (P < .05). The RTV differences for the control and the experimental group were $-7.25{\pm}1.50Ncm$ and $-7.33{\pm}0.93Ncm$, respectively. Statistical analysis revealed that the RTV differences in both groups were not significantly different from each other (P > .05). Conclusion: Within the limitation of this study, the two-piece zirconia abutment with the titanium component did not show a significant RTV difference of the abutment screw compared to the titanium abutment after dynamic cyclic loading.

FIT OF FIXTURE/ABUTMENT INTERFACE OF INTERNAL CONNECTION IMPLANT SYSTEM (내측연결 임플란트 시스템에서 고정체와 지대주 연결부의 적합에 관한 연구)

  • Lee Heung-Tae;Chung Chae-Heon
    • The Journal of Korean Academy of Prosthodontics
    • /
    • v.42 no.2
    • /
    • pp.192-209
    • /
    • 2004
  • Purpose : The purpose of this study was to evaluate the machining accuracy and consistency of implant/abutment/screw combination or internal connection type. Material and methods: In this study, each two randomly selected internal implant fixtures from ITI, 3i, Avana, Bicon, Friadent, Astra, and Paragon system were used. Each abutment was connected to the implant with 32Ncm torque value using a digital torque controller or tapping. All samples were cross-sectioned with grinder-polisher unit (Omnilap 2000 SBT Inc) after embeded in liquid unsaturated polyester (Epovia, Cray Valley Inc). Then optical microscopic and scanning electron microscopic(SEM) evaluations of the implant-abutment interfaces were conducted to assess quality of fit between the mating components. Results : 1) Generally, the geometry of the internal connection system provided for a precision fit of the implant/abutment into interface. 2) The most precision fit of the implant/abutment interface was provided in the case of Bicon System which has not screw. 3) The fit of the implant/abutment interface was usually good in the case of ITI, 3I and Avana system and the amount of fit of the implant/abutment interface was similar to each other. 4) The fit of the implant/abutment interface was usually good in the case of Friadent, Astra and Paragon system. The case of Astra system with the inclined contacting surface had the most Intimate contact among them. 5) Amount of intimate contact in the abutment screw thread to the mating fixture was larger in assembly with two-piece type which is separated screw from abutment such as Friadent, Astra and Paragon system than in that with one-piece type which is not seperated screw from abutment such as ITI, 3I and Avana system. 6) Amount of contact in the screw and the screw seat of abutment was larger in assembly of Friadent system than in asembly of Astra system of Paragon system. Conclusion: Although a little variation in machining accuracy and consistency was noted in the samples, important features of all internal connection systems were the deep, internal implant-abutment connections which provides intimate contact with the implant walls to resist micro-movement, resulting in a strong stable interface. From the results of this study, further research of the stress distribution according to the design of internal connection system will be required.

THE INFLUENCE OF ABUTMENT SCREW TIGHTENING TIMING AND DLC COATING OF CONICAL CONNECTION IMPLANT SYSTEM (일체형 지대주의 Diamond Like Carbon 표면 처리와 나사 조임 시기가 풀림 현상에 미치는 영향)

  • Kim, Ki-Hong;Koak, Jai-Young;Heo, Seong-Joo
    • The Journal of Korean Academy of Prosthodontics
    • /
    • v.46 no.2
    • /
    • pp.209-216
    • /
    • 2008
  • Statement of problem: It is difficult about assessing the loosening tendency of conical connection type implant after cyclic loading with diamond like carbon coating one-piece abutments, and also about the retightening effect. Purpose: This study was performed to investigate the influence of one-piece abutment screw retightening after $5.0{\times}10^{4}$ cyclic loading and Diamond like coating Material and methods: Thirty two ITI implant were divided to 4 groups. Group 1,3-titanium abutment, group 2,4 - diamond like carbon coated abutment. Group 1,2 - $20.0{\times}10^{4}$ cyclic loading after $5.0{\times}10^{4}$ cyclic loading, Group 3,4- after $20.0{\times}10^{4}$ Cyclic loading. After cyclic loading, periotest values were taken and removal torque values of abutments were measured with a digital torque gauge. Results: 1. The removal torque of group 2 after $5.0{\times}10^{4}$ cyclic loading is slightly greater than the other groups but not significantly higher than others (P>0.05). 2. The final removal torque values after $20.0{\times}10^{4}$ cyclic loading of group 1 is bigger than group 3, and group 2 is bigger than group 4, but not significantly higher (P>0.05). 3. The final removal torque values after $20.0{\times}10^{4}$ cyclic loading of all groups are not significantly different (P>0.05).

FIT OF IMPLANT FRAMEWORKS FABRICATED BY ONE-PIECE CASTING, LASER WELDING, SOLDERING, AND ELECTRIC DISCHARGE MACHINING (일체주조법, 레이저용접법, 납착법, 방전가공법에 의해 제작된 임플란트 보철물의 적합도에 관한 연구)

  • Seol, Young-Hoon;Jeong, Chang-Mo;Jeon, Young-Chan;Kang, Sung-Won
    • The Journal of Korean Academy of Prosthodontics
    • /
    • v.40 no.2
    • /
    • pp.156-171
    • /
    • 2002
  • The purpose of this study was to measure and compare the strains produced by screw-tightening implant frameworks fabricated by aye different fabrication methods; (1) one-piece cast using plastic sleeve, (2) one-piece cast using gold cylinder, (3) laser welding, (4) soldering, and (5) electrical discharge machining, and also to measure and compare the strains produced when the order of screw tightening was changed A research model incorporating eighteen strain gages was made to measure the fit of implant frameworks in three dimensions. Three implants aligned in an arc were fixed on the top ends of the L-shape aluminum bars of the research model, and standard abutments were joined to the implants with abutment screws. Five types of implant framework were placed on the abutments and screwed by a torque wrench using 10 Ncm. Under the conditions of this study, the following conclusions were drawn: 1. The electrical discharge machining group showed the smallest magnitude of strain, followed by the soldering group, the laser welding group, the one-piece cast group using gold cylinder, and the one-piece cast group using plastic sleeve. However, among the magnitude of strain for the remaining groups except the electrical discharge machining group, there were not significant differences. 2. When the order of screw tightening was changed, there were not significant differences in the magnitude of strain. 3. In comparison with the electrical discharge machining group, the laser welding group and the one-piece cast groups showed greater horizontal distortion and the soldering group showed greater horizontal and vertical distortion.

Influence of Implant Fixture-Abutment Connection and Abutment Design on Mechanical Strength (임플란트 고정체-지대주 연결부 및 지대주 디자인이 기계적 강도에 미치는 영향)

  • Chun, Mi-Hyun;Jeong, Chang-Mo;Jeon, Young-Chan;Eom, Tae-Gwan;Yoon, Ji-Hoon
    • Journal of Dental Rehabilitation and Applied Science
    • /
    • v.24 no.3
    • /
    • pp.269-281
    • /
    • 2008
  • Fatigue or overload can result in mechanical problems of implant components. The mechanical strength in the implant system is dependent on several factors, such as screw and fixture diameters, material, and design of the fixture-abutment connection and abutment. In these factors, the last rules the strength and stability of the fixture-abutment assembly. There have been some previous reports on the mechanical strength of the fixture-abutment assembly with the compressive bending test or short-term cyclic loading test. However, it is restrictive to predict the long-term stability of the implant system with them. The purpose of this study was to evaluate the influence of the design of the fixture-abutment connection and abutment on the mechanical strength and failure mode by conducting the endurance limit test as well as the compressive bending strength test. Tests were performed according to a specified test(ISO/FDIS 14801) in 4 fixture-abutment assemblies of the Osstem implant system: an external butt joint with Cemented abutment (group BJT), an external butt joint with Safe abutment (group BJS), an internal conical joint with Solid abutment (group CJO), and an internal conical joint with ComOcta abutment (group CJT). The following conclusions were drawn within the limitation of this study. Compressive bending strengths were decreased in order of group BJS(1392.0N), group CJO(1261.8N), group BJT(1153.2N), and group CJT(1110.2N). There were no significant differences in compressive bending strengths between group BJT and group CJT(P>.05). Endurance limits were decreased in order of group CJO(600N), group CJT(453N), group BJS(360N), and group BJT(300N). 3. Compressive bending strengths were influenced by the connection and abutment design of the implant system, however endurance limits were affected more considerably by the connection design.