• Title/Summary/Keyword: OSL 도트

Search Result 1, Processing Time 0.014 seconds

Measuring Absorbed Dose from Medical X-ray Equipment Using Optically Stimulated Luminescence Dots (광자극선량계의 저에너지 엑스선 특성비교)

  • Jung, Sook Jin;Jin, Gye Hwan
    • Journal of the Korean Society of Radiology
    • /
    • v.12 no.1
    • /
    • pp.79-83
    • /
    • 2018
  • In this paper, we measured and analyzed the dose correction factor, absorbed dose linearity, peak voltage X-ray response, angular dependence. Exposure dose correction factor, absorbed dose linearity, and peak voltage linearity using the medical X-ray generator were all in accordance with IEC-62387-1 (2007). The reference to the dosimetry direction at 0, 30, and 60 degrees relative to baseline radiation exposure was -29% (${\pm}30^{\circ}$) and + 67% (${\pm}60^{\circ}$). The values measured at $30^{\circ}$ were -8% lower than the standard and -18% lower than the standard at $60^{\circ}$. Therefore, the effect of direction should be corrected when using OSL dot dosimeter.