• Title/Summary/Keyword: NDIR (Non Dispersive Infrared absorption)

Search Result 11, Processing Time 0.021 seconds

An implementation of NDIR type $CO_2$ gas sample chamber and measuring hardware for capnograph system in consideration of the time response characteristics (시간응답특성을 고려한 2광원 1센서 방식의 capnograph 시스템용 NDIR식 $CO_2$ 가스 챔버 설계 및 측정 회로의 구현)

  • Park, I.Y.;Lee, I.K.;Lee, S.K.;Kang, K.M.;Kang, S.W.;Cho, J.H.
    • Journal of Sensor Science and Technology
    • /
    • v.10 no.5
    • /
    • pp.279-285
    • /
    • 2001
  • The capnograph system for determining the partial pressure of carbon dioxide in the blood of a patient was developed based on the NDIR(non-dispersive infrared) absorption technology. NDIR gas analyzing method requires an optical absorption chamber and signal processing hardware. In this paper, we have designed and implemented NDIR type $CO_2$ gas chamber in consideration of the time response characteristics and lamp chopping frequency. And we have implemented signal processing hardware using two infrared sources to reduce the thermal background effect. The implemented gas chamber and signal processing hardware were tested in the temperature variation experiment and human expiratory experiment. The results showed that the system could produce a stable output signal and a good $CO_2$ gas concentration curve like a typical capnogram.

  • PDF

Development of a Fast-Response $CO_2$ Analyzer using NDIR Technique and Its Application to SI Engine (비분산 적외선 흡수법을 이용한 고속응답 $CO_2$ 분석기의 제작 및 엔진 적용에 관한 연구)

  • Lee, Jae-Young;Min, Kyoung-Doug
    • Transactions of the Korean Society of Automotive Engineers
    • /
    • v.15 no.6
    • /
    • pp.102-107
    • /
    • 2007
  • A fast response $CO_2$ ($fCO_2$) analyzer for real-time measurement of carbon dioxide concentration during transient states of internal combustion engines has been developed. This analyzer uses non-dispersive infrared absorption (NDIR) technique for measuring $CO_2$ concentration and Kalman filter for removing noise components from output signals. The analyzer has good linearity, repeatability and drift with a response time of 11 ms; it is sufficiently fast to detect $CO_2$ concentration during transient states of internal combustion engines. The $fCO_2$ analyzer was used to measure transient $CO_2$ concentration of exhaust gas of the SI engine with a standard gas analyzer, and the signal of the $fCO_2$ analyzer was compared to that of the standard gas analyzer. The two concentrations were well matched during the steady state, and the $fCO_2$ analyzer could measure the variations of $CO_2$ concentration during the transient state.

Composite Gas Measurement System using NDIR Method (NDIR 방법을 이용한 복합 가스 측정 시스템)

  • Eo, Ik-soo
    • Journal of the Korea Academia-Industrial cooperation Society
    • /
    • v.19 no.3
    • /
    • pp.624-629
    • /
    • 2018
  • The current study was conducted to develop a portable composite gas detector allowing the detection of both $CO_2$ and $CH_4$ gases by means of the Non Dispersive Infra-Red (NDIR) method. The gas detector is configured to radiate infrared waves using infrared lamps, where the wavelength of the infrared light is reduced due to absorption throughout the chamber, and this reduction (absorption) is detected by the absorption detector, before being converted and amplified to a 3.5V~6V electrical signal, providing as accurate a measurement as possible. The conventional singe sensor method measures the relative measurement by absorbing only specified wavelengths of infrared radiation, which in the case of gas detection leads to problems with accuracy due to the lack of a reference sensor when detecting light with a wavelength of only $4.26{\mu}m$. The dual sensor employed in this study provides a comparative measurement between the reference value derived from the wavelength of $3.91{\mu}m$, which is not influenced by other gas sources, and the measurement value derived from the wavelength of $4.26{\mu}m$, in order to reduce the errors and enhance the reliability, thereby allowing low power consumption for portable devices and multi-gas detection for both $CO_2$ and $CH_4$ gases. The portable composite gas detector developed herein provides a measurement rage of 0ppm~5,000ppm for $CO_2$ gas, and 0.5%vol for $CH_4$, which allows the determination of whether the $CO_2$ and $CH_4$ contents in indoor air are less than 1,000ppm or not. The current study established that the composite gas detector can be interlinked with firefighting appliances through portable devices or home automation, and is anticipated to be very effective in fire prevention.

Minute Signal Detection Algorithm for Air-pollution Measurement System with The NDIR Detector (NDIR 검출기를 이용하는 대기오염 측정시스템을 위한 미세신호 검출 알고리즘)

  • Choi, Hun;Kim, Hyon-Ho;Whang, Byoung-Han;Lim, Yong-Seok;Ryu, Geun-Taek;Bae, Hyeon-Deok
    • 전자공학회논문지 IE
    • /
    • v.45 no.3
    • /
    • pp.27-35
    • /
    • 2008
  • In this paper, we propose a minute signal detection algorithm for a development of optical analyzer, using the non-dispersive infrared method with multi gas filter correlation wheel, that can measure various environmental air-pollution materials (CO, SO2, NOx, etc.) in real-time. The MCT(mercury cadmium telluride) sensor can detect minute signals those show and absorption characteristic of each environmental pollution materials. In the proposed method, a corresponding data of each environmental pollution materials can be separated by an external trigger and threshold values in the measured continuous signals.

Design of novel optical cavity for NDIR gas sensor with high output power (고출력 NDIR 가스센서를 위한 새로운 optical cavity 설계)

  • Yu, Seon-Hwa;Yi, Seung-Hwan;Kwon, Kwang-Ho;Min, Nam-Ki
    • Proceedings of the KIEE Conference
    • /
    • 2006.07c
    • /
    • pp.1641-1642
    • /
    • 2006
  • 본 연구에서는 NDIR(non dispersive infrared absorption) 가스센서의 소형화와 저가화를 목표로 2개의 오목 거울을 이용한 새로운 광 공동(optical cavity) 구조를 제안하고, 구조에 따른 광 출력 특성을 시뮬레이션하였다. Optical cavity는 광 경로증가 및 광 집속 구조로 설계하고, 각각의 출력단에서 검출되는 광 출력을 시뮬레이션한 결과 광 집속 구조의 팡 출력이 더 효율적임을 확인하였다. 집광구조에서의 광 출력을 최대로 하기 위해 광의 초점을 찾아보았다. 집광 구조를 사용하지 않는 단순 평행 광이 조사되는 경우, 적외선 센서부에 입사되는 광 출력은 최대 0.024W이다. 그러나 집광구조에서는 optical cavity의 광축에서 센서의 위치를 $15\sim20mm$까지 변화시켜 시뮬레이션한 결과 거리 18.83mm 일 때 조사되는 광 출력이 약 0.153W로 약 7배 증가됨을 확인하였다.

  • PDF

Temperature Dependency of Non-dispersive Infrared Carbon Dioxide Gas Sensor by using Infrared Sensor for Compensation (보상용 적외선 센서를 사용한 비분산 적외선 이산화탄소 센서의 온도특성)

  • Yi, SeungHwan
    • Journal of Sensor Science and Technology
    • /
    • v.25 no.2
    • /
    • pp.124-130
    • /
    • 2016
  • NDIR $CO_2$ gas sensor was built with ASIC implemented thermopile sensor which included temperature sensor and unique elliptical waveguide structures in this paper. The temperature dependency of dual infrared sensor module ($CO_2$ and reference IR sensors) has been characterized and its output voltage characteristics according to the temperature and gas concentration were proposed for the first time. NDIR $CO_2$ gas and reference IR sensors showed linear output voltages according to the variation of ambient temperatures from 243 K to 333 K and their slopes were 14.2 mV/K and 8.8 mV/K, respectively. The output voltages of temperature sensor also presented a linear dependency according to the ambient temperature and could be described with V(T)=-3.191+0.0148T(V). The output voltage ratio between $CO_2$ and reference IR sensors revealed irrelevant to the changes of ambient temperatures and gave a constant value around 1.6255 with standard deviation 0.008 at 0 ppm. The output voltage of $CO_2$ gas sensor at zero ppm $CO_2$ gas consisted of two components; one is caused by the HPB (half pass-band) of IR filter and the other is attributed to the part of $CO_2$ absorption wavelength. The characteristics of output voltages of $CO_2$ gas sensor could be accurately modeled with three parameters which are dependent upon the ambient temperatures and represented small average error less than 1.5% with 5% standard deviation.

Temperature Compensation Algorithm of Nondispersive Infrared (NDIR) Gas Sensor (비분산 적외선 가스센서의 온도보상 알고리즘)

  • Park, Jong-Seon;Yi, Seung-Hwan
    • Journal of the Korean Institute of Gas
    • /
    • v.15 no.4
    • /
    • pp.51-55
    • /
    • 2011
  • This paper describes the temperature compensation algorithm using thermopile detector for nondispersive infrared methane gas sensor. From the output voltage of thermistor that is attached onto the infrared detector, the ambient temperature was extracted. The effects of temperatures on the properties of sensor module (the characteristics of narrow bandpass filter, optical cavity and infrared lamp, and gas absorption coefficient times optical path length) have been introduced in order to implement the temperature compensation algorithm. Even though the measurement error of developed sensor module was in the range of $\pm$ 1,500 ppm, after programming the temperature compensation algorithm, the developed sensor module shows a high accuracy less than +180 ppm error within $20^{\circ}C$ temperature variation.

The Emission Characteristics of Bio-Diesel Fuel in Heavy-Duty Engine (바이오 디젤 적용에 따른 대형엔진의 배출가스 특성)

  • Kim, Sun-Moon;Eom, Myoung-Do;Hong, Ji-Hyung
    • Journal of Korean Society for Atmospheric Environment
    • /
    • v.26 no.5
    • /
    • pp.499-506
    • /
    • 2010
  • Recently, a great deal of attention have been directed to the use of alternative fuels as a means to reduce vehicular emissions. As one of the promising alternative fuels, bio-diesel has advantages of a wide adaptability without retrofit of diesel engine. It is also effective enough to reduce CO, THC, $SO_x$, polycyclic aromatic hydrocarbons (PAHs) and PM. In this study, we investigated the emission characteristics of biofuels between different operating conditions, i.e., engine speed (1,400 rpm and 2,300 rpm), engine load (10% and 100%), bio-diesel blending (BD0, BD5 and BD20), and recirculation (EGR) rate of exhaust gas (0% and 20%). Relative performance of the system was evaluated mainly for the greenhouse gases ($CH_4$, $N_2O$ and $CO_2$). In addition, emission characteristics of ND-13 mode were also tested against both greenhouse gases and other airborne pollutants under emission regulation. The relative composition of bio-diesel has shown fairly clear effects on the emission quantities of CO, THC, and PM emission, although it was not on $NO_x$ and greenhouse gases. EGR rate has shown trade-off characteristics between $NO_x$ and PM.

Development of Fast-Response CO2 Analyzer and Analysis of Engine-out Emission during Cold Start of SI Engine (고속응답 CO2 분석기의 제작 및 이를 이용한 SI엔진에서의 실시간 배기가스 분석에 관한 연구)

  • Song, Hyun-Soo;Park, Kyoung-Seok;Park, Dong-Sun;Min, Kyoung-Doug
    • Transactions of the Korean Society of Mechanical Engineers B
    • /
    • v.33 no.2
    • /
    • pp.133-140
    • /
    • 2009
  • A fast-response $CO_2$ analyzer has been developed for measuring the $CO_2$ concentration during transient condition of a SI engine. The analyzer consists of the non-dispersive infrared absorption method, electrical chopping system and water cooling system. The analyzer has good repeatability, linearity and permissible drift characteristic. Besides, it showed 18ms of a response to measure the $CO_2$ concentration. The fast-response $CO_2$ analyzer was applied to a single cylinder SI engine and the $CO_2$ emission was examined during engine start. Simultaneously, the standard exhaust gas analyzer, which has slow response time, was used for confirming the accuracy of the exhaust gas analysis using the fast-response $CO_2$ analyzer. The developed analyzer showed much faster responsive characteristic than that of a standard analyzer and made cycle by cycle exhaust gas analysis possible. The transient engine operating characteristics will be estimated from the $CO_2$ concentration of engine-out emissions and engine operating variables.

Development of Fast-Response $CO_2$ Analyzer and Analysis of Engine-out Emission during Transient Condition of SI engine (고속응답 $CO_2$ 분석기의 제작 및 이를 이용한 SI 엔진에서의 실시간 배기가스 분석에 관한 연구)

  • Song, Hyun-Soo;Min, Kyoung-Doug
    • Proceedings of the KSME Conference
    • /
    • 2008.11b
    • /
    • pp.3079-3084
    • /
    • 2008
  • A fast response $CO_2$ analyzer has been developed for measuring the $CO_2$ concentration during transient condition of SI engine. The analyzer is based on the non-dispersive infrared absorption technique, electrical chopping system and water cooling system. The analyzer has good repeatability, linearity and permissible drift characteristic. Besides, it has 18ms with a response to measure the $CO_2$ concentration. The fast response $CO_2$ analyzer was applied to single cylinder SI engine and the $CO_2$ emission was examined during engine start. Simultaneously, the standard exhaust gas analyzer, which has slow response time, was used for considering the engine-out $CO_2$ characteristic. The developed analyzer showed much faster responsive characteristic than that of a standard analyzer and made cycle by cycle exhaust gas analysis possible. The transient engine operating characteristics will be estimated and the transient behaviors on engine-out emission and performance will be improved.

  • PDF