• Title/Summary/Keyword: MicroVenT CSI-3000

Search Result 1, Processing Time 0.015 seconds

Comparison of the Ventilatory Volume and Airway Pressures using Oxylator EM-100, MicroVenT CSI-3000, OXY-LIFE II -A manikin study - (산소소생기 Oxylator EM-100, MicroVenT CSI-3000, OXY-LIFE II의 환기량과 기도내압 비교 - 마네킨 연구 -)

  • Shim, Gyu-Sik;Kim, Eun-Mee;Roh, Sang-Gyun
    • Fire Science and Engineering
    • /
    • v.31 no.2
    • /
    • pp.113-118
    • /
    • 2017
  • The purpose of this study was to compare the ventilatory volume and airway pressure using Oxylator EM-100, MicroVenT CSI-3000, OXY-LIFE II. The data were obtained from February 13 in 2017 andanalyzed using the SPSS WIN 18.0 program. The results of theventilatory volume showed the Oxylator EM-100 551.44 ml (${\pm}18.70$), MicroVenT CSI-3000 527.26 ml (${\pm}17.98$), and OXY-LIFE II 369.46 ml (${\pm}12.30$). The airway pressure showed the Oxylator EM-100 $11.89cmH_2O$ (${\pm}.41$), MicroVenT CSI-3000 $11.66cmH_2O$ (${\pm}.34$), and OXY-LIFE II $8.02cmH_2O$ (${\pm}.25$). This study will provide the basic data for an appropriate ventilation method by an oxygen supply device including Oxylator EM-100, MicroVenT CSI-3000, and OXY-LIFE II.