A Study on the Image Quality of Mammography and the Average Glandular Dose (맘모그래피의 화질과 평균유선조직선량에 관한 검토)

  • Lee, In-Ja;Kim, Hak-Sung;Kim, Sung-Soo;Huh, Joon
    • Journal of radiological science and technology
    • /
    • v.25 no.2
    • /
    • pp.47-55
    • /
    • 2002
  • We came to the following conclusion as the results of experiment on the image quality of mammography and the average glandular dose using 4 apparatuses at 3 hospitals in Seoul. 1. Whereas the measurement of half value layer showed no differences among the apparatuses, the measurement by an attenuation curve method showed some differences by 5.9%. There were 9.1% differences in the measurement by aluminum conversion method. 2. The basic density of an automatic exposure control unit must be D = 1.40, but there was no automatic exposure unit adjusted precisely at any hospitals. The unit at the B hospital exceeded the allowable limit by ${\pm}0.15$. 3. In the photographing using an automatic exposure control unit and the management of an automatic film processor using a sensitometer, most automatic film processors were well kept. But in some cases the mean value of a fluctuation coefficient exceeded the allowable limit. There is a need for more cautious management. 4. The image quality of breast phantom photography was affected by the screen/film system among the hospitals. 5. The average glandular dose at a breast of 4.2 cm thickness depended on the tube voltage, In the case of Mo/Mo, it was measured $0.26{\sim}1.39\;mGy$ less than ACR standard 3.0 mGy.

  • PDF

The Effect of Scattering Dose on the Thyroid During Mammography (Mammography시 Thyroid에 미치는 산란선량에 관한 연구)

  • Lee, Mi-Hwa;Dong, Kyung-Rae;Park, Seo-Joo;Whang, Sun-Kwang
    • Journal of the Korean Institute of Electrical and Electronic Material Engineers
    • /
    • v.23 no.10
    • /
    • pp.826-830
    • /
    • 2010
  • This study examined the effect of the scattering dose on the thyroid during a mammography examination. One hundred subjects for a mammography examination were enrolled in this study. The average glandular dose (AGD) and thyroid scattering dose (TSD) were measured. Statistical analysis was carried out using the percentage, t-test and co-variance. The mean radiation exposure to the breast and thyroid was $1.08{\pm}0.16$ and $0.14{\pm}0.04$ mGy, respectively. The percentage TSD to the AGD was 31.19%. There was no difference between the Rt. and Lt., and CC to MLO, and radiation dose to the TSD was 13.78% of the breast. Therefore, the volume of radiation exposure to the thyroid was 54.12% in a single routine mammography examination. These results suggest that the TSD was increased by increasing radiation dose to the breast. A thyroid protector is considered necessary to decrease the level of radiation exposure.

Evaluation of Radiation Dose and Imaging of the QC Program in Mammography MLO View (MLO View의 유방촬영에서 QC 프로그램을 이용한 선량 및 영상 평가)

  • Lee, Seon-Hwa;Kim, Jung-Min;Kweon, Dae-Cheol
    • Journal of radiological science and technology
    • /
    • v.38 no.3
    • /
    • pp.221-228
    • /
    • 2015
  • Purpose: In digital mammography QC program was used for the purpose of reducing low-dose and high-definition images of the radiation dose. Materials and Methods: In digital mammography using a QC phantom according to the average glandular dose in the exposure method MLO view $0^{\circ}C$, $30^{\circ}C$, $45^{\circ}C$, $50^{\circ}C$, $55^{\circ}C$, $70^{\circ}C$, was measured at $90^{\circ}C$ intervals, an image with Hologic QC program to the SNR and CNR was measured to evaluate. Results: The average dose in the MLO view was wired to $90^{\circ}C$ when the maximum was 1.75 mGy, it decreased approximately 6% was measured at $45^{\circ}C$ 1.65 mGy. In addition, 1.67 mGy, manual record, there were an average wired in accordance with the exposure dose and the dose of 1.52 mGy difference in the way auto filter. Image quality evaluation at every angular section SNR 50 ~ 52, shows a slight difference in CNR 11 ~ 12, it was included in the manufacturer's recommended value. Conclusion: The dose was lowest in MLO view $45^{\circ}C$, the difference between SNR and CNR were insignificant. The method of exposure will need a way to reduce the exposure of the patient's body or unnecessary patient by placing a difference in settings in which the characteristics.

Patient Dose in Mammography (유방촬영에서 환자 피폭선량)

  • Shin, Gwi-Soon;Kim, You-Hyun;Kim, Jung-Min;Kim, Chang-Kyun;Yang, Jeong-Hwa;Choi, Jong-Hak
    • Journal of radiological science and technology
    • /
    • v.28 no.4
    • /
    • pp.293-299
    • /
    • 2005
  • In the present investigation, we analyzed the data of 1,318 patients (2,636 images) who underwent mammographic examinations and obtained the distribution of the patient age and compressed breast thickness. We measured also average glandular doses (AGD) as function of compressed breast thickness. In order to obtain the values of AGD, we measured half value layer (HVL) and tube output (mR/mAs) for each kVp and target/filter combination. Entrance surface air kerma (ESAK) was calculated from the tube output as measured for each voltage used under clinical conditions and from the tube loading (mAs). AGD per exposure were calculated by multiplying the ESAK values by the conversion factors tabulated by Dance. We obtained in this study the following conclusions. The mean value of compressed breast thickness for cranio-caudal (CC) view was 35.8mm and that for medio-lateral oblique (MLO) view was 43.3 mm. The mean value of AGD for CC view was 1.55 mGy and that for MLO view was 1.70 mGy. The AGD for MLO view was 0.15 mGy (10%) higher than that for CC view because the thickness for MLO view was on average 4.8 mm higher than that for CC view. The values of AGD increased with increasing compressed brest thickness. The increased AGD value was on average 0.34 mGy per 10 mm in the thickness ranges $10{\sim}80\;mm$, therefore differences between the AGD values of each thickness were relative large. Thus, it is considered to need limited doses for mammography with the upper end of exposure range at several different compressed brest thickness.

  • PDF

Patient exposure doses from medical x-ray examinations in Korea (진단방사선검사에서 환자피폭선량에 관한 연구)

  • Kim, You-Hyun;Choi, Jong-Hak;Kim, Sung-Soo;Oh, Yuw-Han;Lee, Chan-Hyeup;Cho, Pyong-Kon;Kang, Dae-Hyun;Lee, Young-Bae;Kim, Hyung-Chul;Kim, Chel-Min
    • Journal of radiological science and technology
    • /
    • v.28 no.3
    • /
    • pp.241-248
    • /
    • 2005
  • X-ray examinations represent the largest man-made source of radiation exposure for the population. The need for standardization of radiation exposures has been suggested and the guidance levels for various radiographic and radioisotope examinations has been proposed by the International Atomic Energy Aency(IAEA) as a safety standard. In many countries, the situation of medical radiographic exposures in each country should be researched before the appropriate guidance level is established. In this study, measurements of entrance surface dose, dose-area product(DAP), computed tomograghic dose index(CTDI) and mean glandular dose(MGD) were carried out in patients who underwent routine x-ray examinations, fluoroscopy, computed tomograghy and mamography in Korea. These measured quantities were compared with the results from the calculation method in previous study. And we suggested diagnostic reference levels in medical imaging in Korea.