Validation of the mobile wireless digital automatic blood pressure monitor using the cuff pressure oscillometric method, for clinical use and self-management, according to international protocols

  • Yoo, Sooyoung;Baek, Hyunyoung;Doh, Kibbeum;Jeong, Jiyeoun;Ahn, Soyeon;Oh, Il-Young;Kim, Kidong
    • Biomedical Engineering Letters
    • /
    • 제8권4호
    • /
    • pp.399-404
    • /
    • 2018
  • The purpose of this study was to evaluate the accuracy of a mobile wireless digital automatic blood pressure monitor for clinical use and mobile health (mHealth). In this study, a manual sphygmomanometer and a digital blood pressure monitor were tested in 100 participants in a repetitive and sequential manner to measure blood pressure. The guidelines for measurement used the Korea Food & Drug Administration protocol, which reflects international standards, such as the American National Standard Institution/Association for the Advancement of Medical Instrumentation SP 10: 1992 and the British Hypertension Society protocol. Measurements were generally consistent across observers according to the measured mean ${\pm}SD$, which ranged in $0.1{\pm}2.6mmHg$ for systolic blood pressure (SBP) and $0.5{\pm}2.2mmHg$ for diastolic blood pressure (DBP). For the device and the observer, the difference in average blood pressure (mean${\pm}$SD) was $2.3{\pm}4.7mmHg$ for SBP and $2.0{\pm}4.2mmHg$ for DBP. The SBP and DBP measured in this study showed accurate measurements that satisfied all criteria, including an average difference that did not exceed 5 mmHg and a standard deviation that did not exceed 8 mmHg. The mobile wireless digital blood pressure monitor has the potential for clinical use and managing one's own health.