  • This paper aims at the development and clinical evaluation of the wireless gyro-mouse system. The wireless gyro-mouse system is a computer interface with gyro-sensor and wireless communication, for the patients with upper-extremity disabled from the traffic accident or stroke to use the computer software i.e. internet browser. In the development, we focused on, firstly, to make the system wireless for the patients to manipulate the mouse easily even on the bed or wheelchair, secondly, to insert the gyro-sensor into a headband for easy don-and-doff and aesthetic appearance, thirdly, to devise a click switch in case of $C5{\sim}C6$ patients and a head nodding detection in case of C4 patients for sending click message to computer operating system. We performed evaluation experiment for patients with upper-extremities disabled from spinal cord injury. The results show that the displacement error of the cursor position against the target position during linear (vertical/horizontal) movement manipulation decreased with trial number. The click rate per minute also increased with trial number. This indicates the developed wireless gyro-mouse system would be more useful to the patients with repetitive use.