• Title/Summary/Keyword: Interaocular Pressure

Search Result 1, Processing Time 0.015 seconds

Design of a Wireless Intraocular Pressure Sensor Based on MEMS Technology (안압의 비접촉 검출을 위한 MEMS 기반의 센서 설계)

  • Kang, Buung-Joo;Park, Jong-Hoon;Lee, So-Hyun;Kang, Ji-Yoon;Park, Chang-Kun
    • Journal of the Korea Institute of Information and Communication Engineering
    • /
    • v.15 no.4
    • /
    • pp.905-912
    • /
    • 2011
  • Interaocular pressure (IOP) sensor and external coil to detect the resonance frequency of the IOP sensor are designed and implemented using MEMS technology. The IOP sensor is designed using 3-D electromagnetic (EM) simulation. The resonance frequency of IOP sensor needs to be lower than that of the external coil. Additionally, the resonance frequency of the IOP sensor needs to be located near the resonance frequency of the coil to get the sufficient amplitude of phase variation. The frequency where the phase peak appears must be constant according to the distance between the IOP sensor and the external coil. From the measurement results, we demonstrated that the designed IOP sensor has the same resonance frequency with various distances between the IOP sensor and the coil.