Algorithm for Integral Method for Photocurrent Measurement of pH Variations Using Multichannel LAPS (다채널 LAPS용 pH 변화량 검출을 위한 적분 방식의 알고리듬 제안)

  • Bae, Sang-Kon;Lee, Sung-Ha;Kang, Shin-Won;Cho, Jin-Ho
    • Journal of the Institute of Electronics Engineers of Korea SC
    • /
    • v.38 no.6
    • /
    • pp.65-75
    • /
    • 2001
  • We proposed the detection method of pH variations by integrating a photocurrent characteristic curve, instead of finding an inflection point by differentiating it in LAPS system. By a simulation of the performance of the proposed method, we verified that it had 80 and 1000 times higher sensitivity and resolution than a conventional method. Then, with the implemented system based on the simulation results, we measured a pH variation which was given rise to a potential change on the LAPS surface exposed to 2-0.03125[mg/ml] enzyme solutions. As results, we observed that the proposed method has a higher sensitivity and resolution of 3.76-0.08[pH/min] pH variations than 3.79-0.27[pH/min] for conventional method with same samples.