• Title/Summary/Keyword: HeaxPOD

Search Result 1, Processing Time 0.015 seconds

Evaluation of the accuracy of the HexaPOD evo RT system using Non-coplanar beams in lung cancer (폐암환자의 비동일평면 선속 빔 치료 시 HexaPOD evo RT system 의 정확성 평가)

  • Jang, Sewuk;cho, Kangchul;Lee, Sangkyoo;Kim, Jooho;Cho, Jeonghee
    • The Journal of Korean Society for Radiation Therapy
    • /
    • v.27 no.2
    • /
    • pp.115-122
    • /
    • 2015
  • Purpose : The aim of this study, evaluate the accuracy of HeaxPOD evo RT system using the non-coplanar beam. Materials and Methods : 13 treatment plans are used which applied non-coplanar beams and 10 treatment plans which coplanar beams are used. the correction value what adjust to 6D couch is determined by each patient's setup errors only rotation direction. The study executed followings. first, Applying the correction value, measure the point dose and calculate the ${\gamma}$-index(${\gamma}=3%$ / 3 mm, ${\gamma}=2%$ / 2 mm). second, acquire data as previous methods without correction by HexaPOD. Results : For comparing the two results, we find out the more precise applying HexaPOD by point dose 0.2% in coplanar and non-coplanar. in the case of ${\gamma}$-index<1(${\gamma}=3%$ / 3 mm), more precise 2.2% in coplanar and 7% in Non-coplanar. Particularly, ${\gamma}$-index<1(2% / 2 mm) show the difference 9.2% in coplanar and 15.1% non-coplanar between apply HexaPOD and dose not apply HexaPOD. Conclusion : Using the HexaPOD is more precise than without HexaPOD. It suggests that HexaPOD evo RT system is very useful for precise and high dose delivery.

  • PDF