Development of a Flexible Eye Shield Phantom for Artifact-Free Computed Tomography Images

  • Seonghee Kang;Chang Heon Choi;Jung-in Kim;Geum Bong Yu;Jin Dong Cho
    • 한국의학물리학회지:의학물리
    • /
    • 제35권3호
    • /
    • pp.59-64
    • /
    • 2024
  • Purpose: This study aimed to develop a flexible eye shield phantom to acquire artifact-free computed tomography (CT) images for electron beam radiotherapy. Methods: A flexible eye shield phantom for a newly designed eye shield was fabricated. Because of metal artifacts caused by an eye shield composed of high-density materials such as tungsten or lead, CT image acquisition is not appropriate for treatment planning because of inaccurate dose calculation and organ-at-risk delineation. To acquire artifact-free CT images, a mold of the same size as the outer dimension of the metallic eye shield was manufactured using 3D printing. The flexible eye shield phantom was imaged using a Philips Brilliance CT Big Bore under the same condition as the measurement. The phantom image with an average of 200 Hounsfield unit (HU) was imported into the treatment planning systems (TPS) and assigned a value of 26,750 HU to consider the material density of tungsten. The dosimetric comparison using a 6-MeV electron beam was performed. Measurement was performed using a metal oxide semiconductor field effect transistor detector for point doses at 3 and 10 mm. Results: The artifact-free CT images using a flexible eye shield phantom without air bubbles were transferred into the TPS. The dose at 10 mm calculated using the TPS agreed with the ion-chamber measurements within 2 cGy. Conversely, a larger dose discrepancy between the measured and calculated doses was found at 3 mm depth. Conclusions: The flexible eye shield phantom was successfully fabricated to apply electron treatment planning by acquiring artifact-free CT images. The dose calculated using the artifact-free image was comparable to the measured dose at lens depth when applying an eye shield.

휘어진 경계에서의 좁은 영역에 대한 Radiochromic 필름 도시메트리 평가 (Evaluation of the Radiochromic Film Dosimetry for a Small Curved Interface)

  • 강세권;박소아;황태진;정광호;한태진;김해영;이미연;김경주;배훈식
    • 한국의학물리학회지:의학물리
    • /
    • 제23권4호
    • /
    • pp.234-238
    • /
    • 2012
  • 눈꺼풀에 발생한 종양의 치료를 위해서는 종종 고에너지 전자선이 이용되며, 이 경우 환자의 시력 보호를 위해 금속차폐체를 눈꺼풀과 안구 사이에 삽입하고 방사선 치료를 시행한다. 차폐체에 접한 눈꺼풀 안쪽의 방사선량 확인을 위해서는 매우 작은 측정도구가 필요하며, 굽은 경계면의 특성상 유연한 측정도구가 바람직한데, radiochromic 필름 도시메트리는 이 목적에 매우 적합하다. 작으면서도 휘어진 경계면을 따라서 선량을 측정하기 위해, 눈꺼풀 팬텀과 차폐체 사이에 3-mm 폭의 EBT2 필름 띠를 삽입하고, 6MeV의 전자선을 조사 후, 선량분포를 얻었다. 금속차폐체와 동일한 크기로 아크릴 재질의 차폐체를 제작하여, 금속인공영상물이 없는 CT 영상을 얻은 후, 이를 이용하여 몬테칼로 전산모사를 수행하였다. 전산모사에서는 실제 안구차폐체의 재질을 따라 텅스텐, 알루미늄 및 스테인레스 스틸 등의 물질 정보를 이용하였다. 이렇게 얻은 전산모사 결과는 필름 측정과 2.1% 내에서 일치하였다. 밀리미터 크기 정도로 작고 또한 휘어진 영역에서 radiochromic 필름 도시메트리는 취급도 용이할 뿐만 아니라 만족스런 정확도를 보여주고 있다.