• Title/Summary/Keyword: Fault Diagnosis of Ultrasonic Probe

Search Result 1, Processing Time 0.016 seconds

Implementation of Capacitance Measurement Equipment for Fault Diagnosis of Multi-channel Ultrasonic Probe (다중채널 초음파 프로브 고장진단을 위한 커패시턴스 측정 장치 구현)

  • Kang, Bub-Joo;Kim, Yang-soo
    • Journal of the Korea Institute of Information and Communication Engineering
    • /
    • v.20 no.1
    • /
    • pp.175-184
    • /
    • 2016
  • In this paper, we propose the method to measure the capacitances using not LCR meter but capacitance to voltage(C/V) conversion. And we design the analog MUX circuits that convert 192 channels to 6 MUX channels in order to implement the diagnosis of multi-channel ultrasonic probe. This paper derives the conversion function that converts the digital voltage of each MUX channel to the capacitance using the least squares method because the circuit characteristics that convert the voltage of each MUX channel to the capacitance are different. The developed prototype illustrates the performance test results that the measure times are measured by within 4sec and the measure error rates of maximum, minimum, and average values are within 5% in terms of the repeated measurements of all 192 channels.