• Title/Summary/Keyword: Electron linear accelerator

Search Result 111, Processing Time 0.025 seconds

Design and Application of Acrylic Electron Wedge for Improving Dose Inhomogeneities at the Junction of Electron Fields (전자선 조사야 결합부분의 선량분포 개선을 위한 acrylic electron wedge의 제작 및 사용)

  • Kim, Young-Bum;Kwon, Young-Ho;Whang, Woong-Ku;Kim, You-Hyun;Kwon, Soo-Il
    • Journal of radiological science and technology
    • /
    • v.21 no.2
    • /
    • pp.36-42
    • /
    • 1998
  • Treatment of a large diseased area with electron often requires the use of two or more adjoining fields. In such cases, not only electron beam divergence and lateral scattering but also fields overlapping and separation may lead to significant dose inhomogeneities(${\pm}20%$) at the region of junction of fields. In this study, we made Acrylic Electron Wedges to improve dose inhomogeneities(${\pm}5%$) in these junction areas and to apply it to clinical practices. All measurements were made using 6, 9, 12, 16, 20 MeV Electron beams from a linear accelerator for a $10{\times}10\;cm$ field at 100cm of SSD. Adding a 1 mm sheet of acryl gradually from 1 mm to 15 mm acquires central axis depth dose beam profile and isodose curves in water phantom. As a result, for all energies, the practical range was reduced by approximately the same distance according to the acryl insert, e.g. a 1 mm thick acryl insert reduces the practical range by approximately 1 mm. For every mm thickness of acryl inserted, the beam energy was reduced to approximately 0.2 MeV. These effects were almost Independent of beam energy and field size. The use of Acrylic Electron Wedges produced a small increase(less than 3%) in the surface dose and a small increase(less than 1%) in X-ray contamination. For acryl inserts, thickness of 3 mm or greater, the penumbra width increased nearly linear for all energies and isodose curves near the beam edge were nearly parallel with the incident beam direction at the point of penumbra width($35\;mm{\sim}40\;mm$). We decide heel thickness and angle of the wedge at this point. These data provide the information necessary to design Acrylic Electron Wedge which can be used to improve dose uniformity at electron field junctions and it will be effectively applied to clinical practices.

  • PDF

Design and Application of Acrylic Electron Wedge to Improve Dose Inhomogeneities at the Junction of Electron Fields (전자선 조사야 결합부분의 선량분포 개선을 위한 Acrylic Electron Wedge의 제작 및 사용)

  • Kim Young Bum;Kwon Young Ho;Whang Woong Ku;Kim You Hyun
    • The Journal of Korean Society for Radiation Therapy
    • /
    • v.10 no.1
    • /
    • pp.60-68
    • /
    • 1998
  • Treatment of a large diseased area with electron often requires the use of two or more adjoining fields. In such cases, not only electron beam divergence and lateral scattering but also fields overlapping and separation may lead to significant dose inhomogeneities(${\pm}20\%$) at the field junction area. In this study, we made Acrylic Electron Wedges to improve dose homogeneities(${\pm}5\%$) in these junction areas and considered application it to clinical practices. All measurements were made using 6, 9, 12, 16, 20MeV Electron beams from a linear accelerator for a $10{\times}10cm$ field at 100cm SSD. Adding a 1 mm sheet of acryl gradually from 1 mm to 15 mm, We acquired central axis depth dose beam profile and isodose curves in water phantom. As a result, for all energies, the practical range was reduced by approximately the same distance as the thickness of the acryl insert, e.g. a 1 mm thick acryl insert reduce the practical range by approximately 1 mm. For every mm thickness of acryl inserted, the beam energy was reduced by approximately 0.2MeV. These effects were almost independent of beam energy and field size. The use of Acrylic Electron Wedges produced a small increase $(less\;than\;3\%)\;in\;the\;surface\;dose\;and\;a\;small\;Increase(less\;than\;1\%)$ in X-ray contamination. For acryl inserts, thickness of 3 mm or greater, the penumbra width increased nearly linear for all energies and isodose curves near the beam edge were nearly parallel with the incident beam direction, and penumbra width was $35\;mm{\sim}40\;mm$. We decide heel thickness and angle of the wedge at this point. These data provide the information necessary to design Acrylic Electron Wedge which can be use to improve dose uniformity at electron field junctions and it will be effectively applicated in clinical practices.

  • PDF

Determination of Initial Beam Parameters of Varian 2100 CD Linac for Various Therapeutic Electrons Using PRIMO

  • Maskani, Reza;Tahmasebibirgani, Mohammad Javad;Hoseini-Ghahfarokhi, Mojtaba;Fatahiasl, Jafar
    • Asian Pacific Journal of Cancer Prevention
    • /
    • v.16 no.17
    • /
    • pp.7795-7801
    • /
    • 2015
  • The aim of the present research was to establish primary characteristics of electron beams for a Varian 2100C/D linear accelerator with recently developed PRIMO Monte Carlo software and to verify relations between electron energy and dose distribution. To maintain conformity of simulated and measured dose curves within 1%/1mm, mean energy, Full Width at Half Maximum (FWHM) of energy and focal spot FWHM of initial beam were changed iteratively. Mean and most probable energies were extracted from validated phase spaces and compared with related empirical equation results. To explain the importance of correct estimation of primary energy on a clinical case, computed tomography images of a thorax phantom were imported in PRIMO. Dose distributions and dose volume histogram (DVH) curves were compared between validated and artificial cases with overestimated energy. Initial mean energies were obtained of 6.68, 9.73, 13.2 and 16.4 MeV for 6, 9, 12 and 15 nominal energies, respectively. Energy FWHM reduced with increase in energy. Three mm focal spot FWHM for 9 MeV and 4 mm for other energies made proper matches of simulated and measured profiles. In addition, the maximum difference of calculated mean electrons energy at the phantom surface with empirical equation was 2.2 percent. Finally, clear differences in DVH curves of validated and artificial energy were observed as heterogeneity indexes were 0.15 for 7.21 MeV and 0.25 for 6.68 MeV. The Monte Carlo model presented in PRIMO for Varian 2100 CD was precisely validated. IAEA polynomial equations estimated mean energy more accurately than a known linear one. Small displacement of R50 changed DVH curves and homogeneity indexes. PRIMO is a user-friendly software which has suitable capabilities to calculate dose distribution in water phantoms or computerized tomographic volumes accurately.

Induced Activity and Space Dose Distribution from Medical Linear Accelerator (의료용(醫療用) 선형가속기(線型加速器)에 의한 산난공간(散亂空間) 선량분포(線量分布)와 유도방사능(誘導放射能))

  • Chu, Sung-Sil;Park, Chang-Yun
    • Journal of Radiation Protection and Research
    • /
    • v.11 no.1
    • /
    • pp.51-56
    • /
    • 1986
  • It is important to measure and protect from the radiation space dose and induced activity at the high energy medical linear accelerator facilities. These are to consider the additional risk to patients undergoing treatment, machine operators and staff members. Measurements of the space dose distribution and induced radioactivity at the 18 MeV medical linear accelerator facility in the Yonsei Cancer Center. 1. Exposure space dose for 300 rads monitor doses of 18 MeV electron are measured as 50 mR at 1 meter from patients. 2. Exposure space dose for 300 rads monitor doses of 10 MV X-ray are detected as 350 mR at 1 meter from phantom. 3. Induced radioactivity by photonuclear reaction was measured as 0.65 mR/hr from collimater after 30 Gy(3,000 rads) irradiated. 4. Analyzing the decay curves and energy spectrum of induced radioactivity, detected a few materials to be activated by photoneutron reaction, $^{65}Cu({\gamma}{\cdot}n)\;^{64}Cu,\;^{186}W({\gamma}{\cdot}n)\;^{185}W,\;^{181}Ta({\gamma}{\cdot}n)\;^{180}Ta,\;^{199}Au({\gamma}{\cdot}n)\;^{198}Au$.

  • PDF

Measurement of Electron Beam Output for the Prototype Compact Linac (콤팩트 전자 선형가속기 시작품의 출력측정에 관한 연구)

  • Kim, Sung-Woo;Kang, Sang Koo;Rhee, Dong Joo;Lim, Heuijin;Lee, Manwoo;Yi, Jungyu;Lee, Mujin;Yang, Kwangmo;Ro, Tae Ik;Jeong, Dong Hyeok
    • Progress in Medical Physics
    • /
    • v.26 no.1
    • /
    • pp.1-5
    • /
    • 2015
  • The C-band compact linear accelerator (linac) is being developed at Dongnam Institute of radiological & Medical Sciences (DIRAMS) for medical and industrial applications. This paper was focused on the output measurement of the electron beam generated from the prototype electron linac. The dose rate was measured in unit of cGy/min per unit pulse frequency according to the IAEA TRS-398 protocol. Exradin-A10 Markus type plane parallel chamber used for the measurement was calibrated in terms of dose to water at the reference depth in water. The beam quality index ($R_{50}$) was determined by the radiochromic film with a solid water phantom approximately due to low energy electrons. As a result, the determined electron beam output was $17.0cGy/(min{\cdot}Hz$. The results were used to monitor the accelerator performance during the development procedure.

Studies on the Interaction of High Energy Electron with Various Matters (물질을 투과한 고에너지 전자선의 선량변화)

  • Chu, S.S.;Kim, G.E.;Park, C.Y.
    • Radiation Oncology Journal
    • /
    • v.1 no.1
    • /
    • pp.11-19
    • /
    • 1983
  • Interaction between high energyelectrons and matters had many complex reactions and the high energy electrons lost their energies with collision and scattering, therefore, electrons distribution in matters was shown as various situation by scattering, exciting and ionizing with moleculars. We experimentally studies with 13 MeV Linear Accelerator and thermoluminescence dosimeter using aluminium and Teflon, etc., and measured energy loss of electrons, electron range, electron scattering and dose distribution in matter. We compared the results with theoretical formular, between 4-qw MeV, the energy loss of electrons was decreased by 2 MeV per $1g/cm^2$ but under 1MeV it was rapidly decreased. Electron range in matter reached to $0.5/cm^2$ per 1MeV of incident energy at 6-12MeV. The dose distribution in matter was increased slightly to some depth by total distribution i.e., the combined intensity of primary and secondary radiant and it was rapidly decreased near the maximum range of electrons. Energy loss of electrons and electron range measured by experiment were coincided with theoretical equations of L. Landau and Feather under 5 and 3% errors respectively. The dose distribution of electrons in matter was similar to L.V. Spencer formular, however, we had found that it was quite different in accordance with the field size and that new formular of dose distribution was induced as empirical function contained experimental factors according to field size.

  • PDF

Fabrication and Characterization of a Fiber-Optic Radiation Sensor for High Energy Electron Beam Therapy (치료용 고에너지 전자선 계측을 위한 광섬유 방사선 센서의 제작 및 특성 분석)

  • Jang, Kyoung-Won;Cho, Dong-Hyun;Yoo, Wook-Jae;Lee, Bong-Soo;Yi, Jeong-Han;Tack, Gye-Rae;Cho, Hyo-Sung;Kim, Sin
    • Journal of Biomedical Engineering Research
    • /
    • v.27 no.6
    • /
    • pp.332-336
    • /
    • 2006
  • In this study, we have fabricated a fiber-optic radiation sensor using an organic scintillator for high energy electron beam therapy. The intensities of scintillating light from a fiber-optic radiation sensor are measured with different field size, electron beam energy and monitor unit of a clinical linear accelerator. To obtain percent depth dose(PDD), the amount of scintillating light is measured at different depth of polymethylmethacrylate(PMMA) phantom. Also the intensity of Cerenkov light is measured and characterized as a function of incident angle of electron beam and a subtraction method is investigated using a background optical fiber to remove a Cerenkov light.

Comparison of Quality of Bologna Sausage Manufactured by Electron Beam or X-Ray Irradiated Ground Pork

  • Shin, Mee-Hye;Lee, Ju-Woon;Yoon, Young-Min;Kim, Jong Heon;Moon, Byeong-Geum;Kim, Jae-Hun;Song, Beom-Suk
    • Food Science of Animal Resources
    • /
    • v.34 no.4
    • /
    • pp.464-471
    • /
    • 2014
  • Ground lean pork was irradiated by an electron beam or X-rays to compare the effects of two types of radiation generated by a linear accelerator on the quality of Bologna sausage as a model meat product. Raw ground lean pork was vacuum packaged at a thickness of 1.5 cm and irradiated at doses of 2, 4, 6, 8, or 10 kGy by an electron beam (2.5 MeV) or X-rays (5 MeV). Solubility of myofibrillar proteins, bacterial counts, and thiobarbituric acid reactive substance (TBARS) values were determined for raw meat samples. Bologna sausage was manufactured using the irradiated lean pork, and total bacterial counts, TBARS values, and quality properties (color differences, cooking yield, texture, and palatability) were determined. Irradiation increased the solubility of myofibrillar proteins in a dose-dependent manner (p<0.05). Bacterial contamination of the raw meat was reduced as the absorbed dose increased, and the reduction was the same for both radiation types. Differences were observed only between irradiated and non-irradiated samples (p<0.05). X-ray irradiation may serve as an alternative to gamma irradiation and electron beam irradiation.

Improvement of Beam-Quality Evaluation Method for Medical Linear Accelerator Using Magnetic Field

  • Kim, Jeongho;Han, Manseok;Yoo, Sejong;Kim, Kijin;Cho, Jae-Hwan
    • Journal of Magnetics
    • /
    • v.20 no.2
    • /
    • pp.120-128
    • /
    • 2015
  • Beam-quality of medical linac evaluations vary by diverse factors. Because conventional beam-quality evaluation methods yield fragmentary results, a new beam-evaluation method is suggested, and its feasibility is evaluated. The PDDs (percentage depth doses) of 6 MV (Mega-voltage) and 10 MV photon, R (Range) of a 6 MeV (Mega Electron-voltage) and 9 MeV electron were measured and compared with the conventional evaluation methods, and the improved methods $PDD^{10}{_5}$, $PDD^{20}{_{10}}$, $PDD^{30}{_{20}}$, $PDD^{20}{_5}$, $PDD^{30}{_{10}}$, and $R^{70}{_{50}}$, $R^{50}{_{30}}$, $R^{70}{_{30}}$ as the magnetic field of the bending magnet was changed to +2% to -2%, and the results were compared. The comparison showed that the improved methods exhibit a higher discrimination than the conventional methods in each energy regime. $PDD^{10}{_5}$, $PDD^{30}{_{20}}$, $PDD^{30}{_{10}}$ and $R^{70}{_{50}}$, $R^{50}{_{30}}$ should be applied. These methods exhibit a higher discrimination in each energy regime than conventional beam-quality evaluation methods; therefore, they should be used for beam-quality evaluation according to the magnetic field variation.

ML-6M선형 가속기에서의 BEAM 특성에 대한 고찰

  • Mun, Eon-Cheol;Yun, Byeong-Un;O, Yang-Hwan
    • The Journal of Korean Society for Radiation Therapy
    • /
    • v.5 no.1
    • /
    • pp.115-119
    • /
    • 1992
  • The beam characteristics and dosimetric measurements of the 6MV X-ray and 6MeV electron beam from a ML-6M linear accelerator are examined. The Percent Depth Dose(PDD) table and the tissue Maximum Ratio(TMR) table are taken from measurement as a function of the field size and the depth. The calculated TMR table from PDD table is compared with those from measurement. Other beam characteristics such as output factor, beam profile(including flatness, symmetry and penumbra), wedge, and the variation of Dmax are presented. All of these dosimetric measurements sufficiently characterized the beam to permit safe clinical use.

  • PDF