Effects of Tube Voltage and Tube Current on Exposure Index : Focused on Radiographic Images of Cone Pyramid Phantom (관전압과 관전류량이 노출 지수에 미치는 영향 : 원뿔형 피라미드 팬텀 방사선영상 중심으로)

  • Seoung, You-Hun
    • Journal of the Korean Society of Radiology
    • /
    • v.13 no.5
    • /
    • pp.749-755
    • /
    • 2019
  • In this study, we investigated the effects of tube voltage (kVp) and tube current (mAs) on an exposure index (EI) applied to a digital radiography. There used an inverter type digital X-ray generator and an image receptor (IR) utilized a portable wireless detector without a grid. Radiographic images were acquired using a cone pyramid phantom produced using a 3D printer. The X-ray tube voltage was increased from 40 kVp to 120 kVp with 10 kVp increment, and the tube current was increased from 1 mAs to 128 mAs with the twice increment. As a result, kVp had logarithmic relationship with the EI as high $R^2$ value, and mAs had a very high linear relationship too. Also, there was a high correlation between the area dose and the EI of the IR, with the $R^2$ value being 0.76 or more. In conclusion, it had showed that mAs affected the EI linearly, and that it could be advantageous to adjust the easy-to-predict mAs to maintain proper image qualities.