• Title/Summary/Keyword: Continuous Preconcentration-HG

Search Result 1, Processing Time 0.014 seconds

Continuous Coprecipitantion Preconcentration-Hydride Generation for Arsenic in Inductively Coupled Plasma-Atomic Emission Spectrometry (연속적 공침 선농축-수소화물 발생법을 이용한 ICP-AES에서의 비소의 감도 개선)

  • Kim, Chang-Gyu;Pak, Yong-Nam
    • Journal of the Korean Chemical Society
    • /
    • v.48 no.6
    • /
    • pp.583-589
    • /
    • 2004
  • In a stream of aqueous sample, trace arsenic ions were quantitatively coprecipitated and detected in ICP-AES through hydride generation. In was used as a coprecipitating reagent. The precipitate was collected on a filter and dissolved by HCl. The eluted As was sent into the reaction coil to generate hydrides and analyzed by ICP. With optimal conditions, and with a sample of 0.3 mL, an enrichment of 70 was obtained with the sampling speed of 10/hr. When compared with coprecipitation and hydride generation technique, the sensitivity was increased by 7 and 10 times, respectively. The limit of detection limit$(3{\sigma})$ was 0.020 ${\mu}g\;L^{-1}$ and the precision was 7-10%. Separation of $As^{3+}\;and\;As^{5+}$ were possible using citric acid in hydride generation.