• Title/Summary/Keyword: Calibration of Neutron Survey Meter

Search Result 4, Processing Time 0.017 seconds

A REVIEW OF NEUTRON SCATTERING CORRECTION FOR THE CALIBRATION OF NEUTRON SURVEY METERS USING THE SHADOW CONE METHOD

  • KIM, SANG IN;KIM, BONG HWAN;KIM, JANG LYUL;LEE, JUNG IL
    • Nuclear Engineering and Technology
    • /
    • v.47 no.7
    • /
    • pp.939-944
    • /
    • 2015
  • The calibration methods of neutron-measuring devices such as the neutron survey meter have advantages and disadvantages. To compare the calibration factors obtained by the shadow cone method and semi-empirical method, 10 neutron survey meters of five different types were used in this study. This experiment was performed at the Korea Atomic Energy Research Institute (KAERI; Daejeon, South Korea), and the calibration neutron fields were constructed using a $^{252}Californium$ ($^{252}Cf$) neutron source, which was positioned in the center of the neutron irradiation room. The neutron spectra of the calibration neutron fields were measured by a europium-activated lithium iodide scintillator in combination with KAERI's Bonner sphere system. When the shadow cone method was used, 10 single moderator-based survey meters exhibited a smaller calibration factor by as much as 3.1-9.3% than that of the semi-empirical method. This finding indicates that neutron survey meters underestimated the scattered neutrons and attenuated neutrons (i.e., the total scatter corrections). This underestimation of the calibration factor was attributed to the fact that single moderator-based survey meters have an under-ambient dose equivalent response in the thermal or thermal-dominant neutron field. As a result, when the shadow cone method is used for a single moderator-based survey meter, an additional correction and the International Organization for Standardization standard 8529-2 for room-scattered neutrons should be considered.

Determination of Microdosimetric Quantities of Several Neutron Calibration Fields at KAERI

  • Kim, B.H.;Kim, J.S.;Kim, J.L.;Chang, S.Y.;Cho, G.;McDonald, J.C.
    • Journal of Radiation Protection and Research
    • /
    • v.28 no.4
    • /
    • pp.327-335
    • /
    • 2003
  • The commercially available neutron survey meter, the REM500, which uses a tissue equivalent proportional counter (TEPC) and the self-constructed TEPC were used to determine the microdosimetric quantities of several neutron calibration fields at Korea Atomic Energy Research Institute (KAERI). Microdosimetric spectra, absorbed dose, dose equivalent as well as quality factor were derived and compared with several neutron fields which were produced by using the shadow objects to make neutron scattered and being used as a kind of realistic neutron calibration fields at KAERI. The response of REM500 as a function of mean energy was evaluated with these neutron fields using the counts measured and the predetermined reference value. The response of the self-made TEPC and the REM500 was compared using one of the neutron calibration filelds of a $^{252}Cf$ source. The reference quantities of scattered neutron calibration fields were determined using a Bonner Sphere (BS). The value of frequency-mean lineal energy, dose-mean lineal energy and quality factor of two $^{252}Cf$ sources (unmoderated and $D_2O$ moderated) were determined to check the differences in the reference neutron fields between KAERI and Pacific Northwest National Laboratory (PNNL, USA) and the results were in good agreement within 1%. It means that there is no big difference in dosimetric quantifies of neutron calibration fields of two laboratories.

SPECTRUM WEIGHTED RESPONSES OF SEVERAL DETECTORS IN MIXED FIELDS OF FAST AND THERMAL NEUTRONS

  • Kim, Sang In;Chang, Insu;Kim, Bong Hwan;Kim, Jang Lyul;Lee, Jung Il
    • Nuclear Engineering and Technology
    • /
    • v.46 no.2
    • /
    • pp.273-280
    • /
    • 2014
  • The spectrum weighted responses of various detectors were calculated to provide guidance on the proper selection and use of survey instruments on the basis of their energy response characteristics on the neutron fields. To yield the spectrum weighted response, the detector response functions of 17 neutron-measuring devices were numerically folded with each of the produced calibration neutron spectra through the in-house developed software 'K-SWR'. The detectors' response functions were taken from the IAEA Technical Reports Series No. 403 (TRS-403). The reference neutron fields of 21 kinds with 2 spectra groups with different proportions of thermal and fast neutrons have been produced using neutrons from the $^{241}Am$-Be sources held in a graphite pile, a bare $^{241}Am$-Be source, and a DT neutron generator. Fluence-average energy ($E_{ave}$) varied from 3.8 MeV to 16.9 MeV, and the ambient-dose-equivalent rate [$H^*(10)/h$] varied from 0.99 to 16.5 mSv/h.

Development of Neutron Induced Prompt γ-ray Spectroscopy System Using 252Cf (252Cf 선원을 이용한 즉발감마선 계측시스템 구성)

  • Park, Yong-Joon;Song, Byung-Chul;Jee, Kwang-Yong
    • Analytical Science and Technology
    • /
    • v.16 no.1
    • /
    • pp.12-24
    • /
    • 2003
  • For the design and set-up of neutron induced prompt ${\gamma}$-ray spectroscopy system using $^{252}Cf$ neutron source, the effects of shielding and moderator materials have been examined. The $^{252}Cf$ source being used for TLD badge calibration in Korea Atomic Energy Research Institute was utilized for this preliminary experiment. The ${\gamma}$-ray background and prompt ${\gamma}$-ray spectrum of the sample containing Cl were measured using HPGe (GMX 60% relative efficiency) located at the inside of the system connected to notebook PC at the outside of the system (about 20 meter distance). The background activities of neutron and ${\gamma}$-rays were measured with neutron survey meter as well as ${\gamma}$-ray survey meters, respectively and the system was designed to minimize the activities. Prompt ${\gamma}$-ray spectrum was measured using ${\gamma}$-${\gamma}$ coincident system for reduce the background and the continuum spectrum. The optimum system was designed and set up using the experimental data obtained.