• Title/Summary/Keyword: CZT Spear Detector

Search Result 1, Processing Time 0.015 seconds

Development of One-channel Gamma ray spectroscope for Automatic Radiopharmaceutical Synthesis System (방사성 의약품 자동합성장치용 단채널 감마선 분광기 보드의 설계 및 제작)

  • Song, Kwanhoon;Kim, Kwangsoo
    • Journal of the Institute of Electronics and Information Engineers
    • /
    • v.51 no.4
    • /
    • pp.193-200
    • /
    • 2014
  • In this paper, the prototype of one-channel gamma-ray spectroscope for automatic radiopharmaceutical systhesis system was designed and characterized. The prototype employed CZT (CdZnTe) spear detector for gamma-ray detection and employed analog-type signal processing method. A radioactive sample Co-60 was used for measuring performance of the gamma-ray spectroscope and energy spectrum is gained with bandwidth of 1173keV. The analog board is made up of SF (shaping filter) and PHA (peak and hold amplifier) for shaping CZT output signal appropriately and ADC (analog to digital converter) and FPGA (field programmable gate array) for drawing gamma-ray spectrum by counting the digitalized gamma-ray signal data.