• Title/Summary/Keyword: CT장치

Search Result 316, Processing Time 0.026 seconds

Evaluation of Ovary Dose for woman of Childbearing age Woman with Breast cancer in tomotherapy (가임기 여성의 유방암 토모치료 시 난소선량 평가비교)

  • Lee, Soo Hyeung;Park, Soo Yeun;Choi, Ji Min;Park, Ju Young;Kim, Jong Suk
    • The Journal of Korean Society for Radiation Therapy
    • /
    • v.26 no.2
    • /
    • pp.337-343
    • /
    • 2014
  • Purpose : The aim of this study is to evaluate unwanted scattered dose to ovary by scattering and leakage generated from treatment fields of Tomotherapy for childbearing woman with breast cancer. Materials and Methods : The radiation treatments plans for left breast cancer were established using Tomotherapy planning system (Tomotherapy, Inc, USA). They were generated by using helical and direct Tomotherapy methods for comparison. The CT images for the planning were scanned with 2.5 mm slice thickness using anthropomorphic phantom (Alderson-Rando phantom, The Phantom Laboratory, USA). The measurement points for the ovary dose were determined at the points laterally 30 cm apart from mid-point of treatment field of the pelvis. The measurements were repeated five times and averaged using glass dosimeters (1.5 mm diameter and 12 mm of length) equipped with low-energy correction filter. The measures dose values were also converted to Organ Equivalent Dose (OED) by the linear exponential dose-response model. Results : Scattered doses of ovary which were measured based on two methods of Tomo helical and Tomo direct showed average of $64.94{\pm}0.84mGy$ and $37.64{\pm}1.20mGy$ in left ovary part and average of $64.38{\pm}1.85mGy$ and $32.96{\pm}1.11mGy$ in right ovary part. This showed when executing Tomotherapy, measured scattered dose of Tomo Helical method which has relatively greater monitor units (MUs) and longer irradiation time are approximately 1.8 times higher than Tomo direct method. Conclusion : Scattered dose of left and right ovary of childbearing women is lower than ICRP recommended does which is not seriously worried level against the infertility and secondary cancer occurrence. However, as breast cancer occurrence ages become younger in the future and radiation therapy using high-precision image guidance equipment like Tomotherapy is developed, clinical follow-up studies about the ovary dose of childbearing women patients would be more required.

Feasibility of MatriXX for Intensity Modulated Radiation Therapy Quality Assurance (세기변조방사선치료의 품질관리를 위한 이온전리함 매트릭스의 유용성 고찰)

  • Kang, Min-Young;Kim, Yoen-Lae;Park, Byung-Moon;Bae, Yong-Ki;Bang, Dong-Wan
    • The Journal of Korean Society for Radiation Therapy
    • /
    • v.19 no.2
    • /
    • pp.91-97
    • /
    • 2007
  • Purpose: To evaluate the feasibility of a commercial ion chamber array for intensity modulated radiation therapy (IMRT) quality assurance (QA) was performed IMRT patient-specific QA Materials and Methods: A use of IMRT patient-specific QA was examined for nasopharyngeal patient by using 6MV photon beams. The MatriXX (Wellhofer Dosimetrie, Germany) was used for IMRT QA. The case of nasopharyngeal cancer was performed inverse treatment planning. A hybrid dose distribution made on the CT data of MatriXX and solid phantom all of the same gantry angle (0$^\circ$). The measurement was acquired with geometrical condition that equal to hybrid treatment planning. The $\gamma$-index (dose difference 3%, DTA 3 mm) histogram was used for quantitative analysis of dose discrepancies. An absolute dose was compared at the high dose low gradient region. Results: The dose distribution was shown a good agreement by gamma evaluation. A proportion of acceptance criteria was 95.8%, 97.52%, 96.28%, 98.20%, 97.78%, 96.64% and 92.70% for gantry angles were 0$^\circ$, 55$^\circ$, 110$^\circ$, 140$^\circ$, 220$^\circ$, 250$^\circ$ and 305$^\circ$, respectively. The absolute dose in high dose low gradient region was shown reasonable agreement with the RTP calculation within $\pm$3%. Conclusion: The MatriXX offers the dosimetric characteristics required for performing both relative and absolute measurements. If MatriXX use in the clinic, it could be simplified and reduced the IMRT patient-specific QA workload. Therefore, the MatriXX is evaluated as a reliable and convenient dosimeter for IMRT patient-specific QA.

  • PDF

Patients Setup Verification Tool for RT (PSVTS) : DRR, Simulation, Portal and Digital images (방사선치료 시 환자자세 검증을 위한 분석용 도구 개발)

  • Lee Suk;Seong Jinsil;Kwon Soo I1;Chu Sung Sil;Lee Chang Geol;Suh Chang Ok
    • Radiation Oncology Journal
    • /
    • v.21 no.1
    • /
    • pp.100-106
    • /
    • 2003
  • Purpose : To develop a patients' setup verification tool (PSVT) to verify the alignment of the machine and the target isocenters, and the reproduclbility of patients' setup for three dimensional conformal radiotherapy (3DCRT) and intensity modulated radiotherapy (IMRT). The utilization of this system is evaluated through phantom and patient case studies. Materials and methods : We developed and clinically tested a new method for patients' setup verification, using digitally reconstructed radiography (DRR), simulation, porial and digital images. The PSVT system was networked to a Pentium PC for the transmission of the acquired images to the PC for analysis. To verify the alignment of the machine and target isocenters, orthogonal pairs of simulation images were used as verification images. Errors in the isocenter alignment were measured by comparing the verification images with DRR of CT Images. Orthogonal films were taken of all the patients once a week. These verification films were compared with the DRR were used for the treatment setup. By performing this procedure every treatment, using humanoid phantom and patient cases, the errors of localization can be analyzed, with adjustments made from the translation. The reproducibility of the patients' setup was verified using portal and digital images. Results : The PSVT system was developed to verify the alignment of the machine and the target isocenters, and the reproducibility of the patients' setup for 3DCRT and IMRT. The results show that the localization errors are 0.8$\pm$0.2 mm (AP) and 1.0$\pm$0.3 mm (Lateral) in the cases relating to the brain and 1.1$\pm$0.5 mm (AP) and 1.0$\pm$0.6 mm (Lateral) in the cases relating to the pelvis. The reproducibility of the patients' setup was verified by visualization, using real-time image acquisition, leading to the practical utilization of our software Conclusions : A PSVT system was developed for the verification of the alignment between machine and the target isocenters, and the reproduclbility of the patients' setup in 3DCRT and IMRT. With adjustment of the completed GUI-based algorithm, and a good quality DRR image, our software may be used for clinical applications.

Evaluation of Set-up Accuracy for Frame-based and Frameless Lung Stereotactic Body Radiation Therapy (폐암 정위체부방사선치료 시 고정기구(frame) 사용 유무에 따른 셋업 정확성 평가)

  • Ji, Yunseo;Chang, Kyung Hwan;Cho, Byungchul;Kwak, Jungwon;Song, Si Yeol;Choi, Eun Kyung;Lee, Sang-wook
    • Progress in Medical Physics
    • /
    • v.26 no.4
    • /
    • pp.286-293
    • /
    • 2015
  • The purpose of this study was to evaluate the set up accuracy using stereotactic body frame and frameless immobilizer for lung stereotactic body radiation therapy (SBRT). For total 40 lung cancer patients treated by SBRT, 20 patients using stereotactic body frame and other 20 patients using frameless immobilizer were separately enrolled in each group. The setup errors of each group depending on the immobilization methods were compared and analyzed. All patients received the dose of 48~60 Gy for 4 or 5 fractions. Before each treatment, a patient was first localized to the treatment isocenter using room lasers, and further aligned with a series of image guidance procedures; orthogonal kV radiographs, cone-beam CT, orthogonal fluoroscopy. The couch shifts during these procedures were recorded and analyzed for systematic and random errors of each group. Student t-test was performed to evaluate significant difference depending on the immobilization methods. The setup reproducibility was further analyzed using F-test with the random errors excluding the systematic setup errors. In addition, the ITV-PTV margin for each group was calculated. The setup errors for SBF were $0.05{\pm}0.25cm$ in vertical direction, $0.20{\pm}0.38cm$ in longitudinal direction, and $0.02{\pm}0.30cm$ in lateral direction, respectively. However the setup errors for frameless immobilizer showed a significant increase of $-0.24{\pm}0.25cm$ in vertical direction while similar results of $0.06{\pm}0.34cm$, $-0.02{\pm}0.25cm$ in longitudinal and lateral directions. ITV-PTV margins for SBF were 0.67 cm (vertical), 0.99 cm (longitudinal), and 0.83 cm (lateral), respectively. On the other hand, ITV-PTV margins for Frameless immobilizer were 0.75 cm (vertical), 0.96 cm (longitudinal), and 0.72 cm (lateral), indicating less than 1 mm difference for all directions. In conclusion, stereotactic body frame improves reproducibility of patient setup, resulted in 0.1~0.2 cm in both vertical and longitudinal directions. However the improvements are not substantial in clinic considering the effort and time consumption required for SBF setup.

Development of Movement Analysis Program and its Feasibility Test in Streotactic Body Radiation Threrapy (복부부위의 체부정위방사선치료시 호흡에 의한 움직임분석 프로그램 개발 및 유용성 평가)

  • Shin, Eun-Hyuk;Han, Young-Yih;Kim, Jin-Sung;Park, Hee-Chul;Shin, Jung-Suk;Ju, Sang-Gyu;Lee, Ji-Hea;Ahn, Jong-Ho;Lee, Jai-Ki;Choi, Doo-Ho
    • Progress in Medical Physics
    • /
    • v.22 no.3
    • /
    • pp.107-116
    • /
    • 2011
  • Respiratory gated radiation therapy and stereotactic body radiation therapy require identical tumor motions during each treatment with the motion detected in treatment planning CT. Therefore, this study developed a tumor motion monitoring and analysis system during the treatments employing RPM data, gated setup OBI images and a data analysis software. A respiratory training and guiding program which improves the regularity of breathing was used to patients. The breathing signal was obtained by RPM and the recorded data in the 4D console was read after treatment. The setup OBI images obtained gated at 0% and 50% of breathing phases were used to detect the tumor motion range in crenio-caudal direction. By matching the RPM data recorded at the OBI imaging time, a factor which converts the RPM motion to the tumor motion was computed. RPM data was entered to the institute developed data analysis software and the maximum, minimum, average of the breathing motion as well as the standard deviation of motion amplitude and period was computed. The computed result is exported in an excel file. The conversion factor was applied to the analyzed data to estimate the tumor motion. The accuracy of the developed method was tested by using a moving phantom, and the efficacy was evaluated for 10 stereotactic body radiation therapy patients. For the sine wave motion of the phantom with 4 sec of period and 2 cm of peak-to-peak amplitude, the measurement was slightly larger (4.052 sec) and the amplitude was smaller (1.952 cm). For patient treatment, one patient was evaluated not to qualified to SBRT due to the usability of the breathing, and in one patient case, the treatment was changed to respiratory gated treatment due the larger motion range of the tumor than treatment planed motion. The developed method and data analysis program was useful to estimate the tumor motion during treatment.

Comparison of Helical TomoTherapy with Linear Accelerator Base Intensity-modulated Radiotherapy for Head & Neck Cases (두경부암 환자에 대한 선량체적 히스토그램에 따른 토모치료외 선형가속기기반 세기변조방사선치료의 정량적 비교)

  • Kim, Dong-Wook;Yoon, Myong-Geun;Park, Sung-Yong;Lee, Se-Byeong;Shin, Dong-Ho;Lee, Doo-Hyeon;Kwak, Jung-Won;Park, So-Ah;Lim, Young-Kyung;Kim, Jin-Sung;Shin, Jung-Wook;Cho, Kwan-Ho
    • Progress in Medical Physics
    • /
    • v.19 no.2
    • /
    • pp.89-94
    • /
    • 2008
  • TomoTherapy has a merit to treat cancer with Intensity modulated radiation and combines precise 3-D imaging from computerized tomography (CT scanning) with highly targeted radiation beams and rotating beamlets. In this paper, we comparing the dose distribution between TomoTherapy and linear accelerator based intensity modulated radiotherapy (IMRT) for 10 Head & Neck patients using TomoTherapy which is newly installed and operated at National Cancer Center since Sept. 2006. Furthermore, we estimate how the homogeneity and Normal Tissue Complication Probability (NTCP) are changed by motion of target. Inverse planning was carried out using CadPlan planning system (CadPlan R.6.4.7, Varian Medical System Inc. 3100 Hansen Way, Palo Alto, CA 94304-1129, USA). For each patient, an inverse IMRT plan was also made using TomoTherapy Hi-Art System (Hi-Art2_2_4 2.2.4.15, TomoTherapy Incorporated, 1240 Deming Way, Madson, WI 53717-1954, USA) and using the same targets and optimization goals. All TomoTherapy plans compared favorably with the IMRT plans regarding sparing of the organs at risk and keeping an equivalent target dose homogeneity. Our results suggest that TomoTherapy is able to reduce the normal tissue complication probability (NTCP) further, keeping a similar target dose homogeneity.

  • PDF

Guideline for Imaging Dose on Image-Guided Radiation Therapy (영상유도방사선치료에 있어 영상선량 가이드라인)

  • Cho, Byung Chul;Huh, Hyun Do;Kim, Jin Sung;Choi, Jin Ho;Kim, Seong Hoon;Cho, Kwang Hwan;Cho, Sam Ju;Min, Chul Kee;Shin, Dong Oh;Lee, Sang Hoon;Park, Dong Wook;Kim, Kum Bae;Choi, Sang Hyoun;Kim, Hye Young;Ahn, Woo-Sang;Kim, Tae Hyeong;Han, Su Cheol
    • Progress in Medical Physics
    • /
    • v.24 no.1
    • /
    • pp.1-24
    • /
    • 2013
  • As image-guided radiation therapy (IGRT) has been commonly used for more accurate patient setup and monitoring tumor movement during radiation therapy, the necessity for management of imaging dose is increased. However, it has not been an interest issue to radiation therapy communities because the imaging dose is much lower than the therapeutic dose. However, since the cumulative dose from 4DCT and repeated imaging for daily setup verificationin would not be ignorable, appropriate dose management based on ALARA (As Low As Reasonably Achievable) principle is required. In this study, we aimed that (1) survey on imaging equipments and modalities used for IGRT, (2) estimation of IGRT imaging dose depending on treatment types and equipments, (3) collecting data of effective dose on treatment sites from each equipment and imaging protocol, and thus finally provide guideline for imaging dose reduction and optimization.

Treatment Margin Assessment using Mega-Voltage Computed Tomography of a Tomotherapy Unit in the Radiotherapy of a Liver Tumor (간종양 방사선치료 시 토모테라피 메가볼트 CT를 이용한 치료 여백 평가)

  • You, Sei-Hwan;Seong, Jin-Sil;Lee, Ik-Jae;Koom, Woong-Sub;Jeon, Byeong-Chul
    • Radiation Oncology Journal
    • /
    • v.26 no.4
    • /
    • pp.280-288
    • /
    • 2008
  • Purpose: To identify the inter-fractional shift pattern and to assess an adequate treatment margin in the radiotherapy of a liver tumor using mega-voltage computed tomography (MVCT) of a tomotherapy unit. Materials and Methods: Twenty-six patients were treated for liver tumors by tomotherapy from April 2006 to August 2007. The MVCT images of each patient were analyzed from the $1^{st}$ to the $10^{th}$ fraction for the assessment of the daily liver shift by four groups based on Couinard's proposal. Daily setup errors were corrected by bony landmarks as a prerequisite. Subsequently, the anterior-, posterior-, right-, and left shifts of the liver edges were measured by maximum linear discrepancies between the kilo-voltage computed tomography (KVCT) image and MVCT image. All data were set in the 2-dimensional right angle coordinate system of the transverse section of each patient's body. Results: The liver boundary shift had different patterns for each group. In group II (segment 2, 3, and 4), the anterior mean shift was $2.80{\pm}1.73\;mm$ outwards, while the left mean shift was $2.23{\pm}1.37\;mm$ inwards. In group IV (segment 7 and 8), the anterior-, posterior-, right-, and left mean shifts were $0.15{\pm}3.93\;mm$ inwards, $3.15{\pm}6.58\;mm$ inwards, $0.60{\pm}3.58\;mm$ inwards, and $4.50{\pm}5.35\;mm$ inwards, respectively. The reduced volume in group II after MVCT reassessment might be a consequence of stomach toxicity. Conclusion: Inter-fractional liver shifts of each group based on Couinard's proposal were somewhat systematic despite certain variations observed in each patient. The geometrical deformation of the liver by respiratory movement can cause shrinkage in the left margins of liver. We recommend a more sophisticated approach in free-breathing mode when irradiating the left lobe of liver in order to avoid stomach toxicity.

A Study of Various Filter Setups with FBP Reconstruction for Digital Breast Tomosynthesis (디지털 유방단층영상합성법의 FBP 알고리즘 적용을 위한 다양한 필터 조합에 대한 연구)

  • Lee, Haeng-Hwa;Kim, Ye-Seul;Lee, Youngjin;Choi, Sunghoon;Lee, Seungwan;Park, Hye-Suk;Kim, Hee-Joung;Choi, Jae-Gu;Choi, Young-Wook
    • Progress in Medical Physics
    • /
    • v.25 no.4
    • /
    • pp.271-280
    • /
    • 2014
  • Recently, digital breast tomosynthesis (DBT) has been investigated to overcome the limitation of conventional mammography for overlapping anatomical structures and high patient dose with cone-beam computed tomography (CBCT). However incomplete sampling due to limited angle leads to interference on the neighboring slices. Many studies have investigated to reduce artifacts such as interference. Moreover, appropriate filters for tomosynthesis have been researched to solve artifacts resulted from incomplete sampling. The primary purpose of this study is finding appropriate filter scheme with FBP reconstruction for DBT system to reduce artifacts. In this study, we investigated characteristics of various filter schemes with simulation and prototype digital breast tomosynthesis under same acquisition parameters and conditions. We evaluated artifacts and noise with profiles and COV (coefficinet of variation) to study characteristic of filter. As a result, the noise with parameter 0.25 of Spectral filter reduced by 10% in comparison to that with only Ramp-lak filter. Because unbalance of information reduced with decreasing B of Slice thickness filter, artifacts caused by incomplete sampling reduced. In conclusion, we confirmed basic characteristics of filter operations and improvement of image quality by appropriate filter scheme. The results of this study can be utilized as base in research and development of DBT system by providing information that is about noise and artifacts depend on various filter schemes.

The variability of 6-D Skull Tracking(6DST) in Cyberknife for Bone metastasis patients (사이버나이프 6-D Skull Tracking의 유용성 평가)

  • Lee, Geon Ho;Bae, Sun Myeong;Song, Heung Kwon;Baek, Geum Mun
    • The Journal of Korean Society for Radiation Therapy
    • /
    • v.30 no.1_2
    • /
    • pp.41-47
    • /
    • 2018
  • Purpose : The purpose of this study is to evaluate the usefulness of 6 Dimensional Skull Tracking(6DST) in Cyberknife Stereotactic Body Radiation Therapy(SBRT) for the first and second cervical vertebrae(C1 and C2) metastasis. Methode and material : The Computed Tomography (Lightspeed VCT 64, General Electric Co. Waukesha, WI, USA) was used to acquire the CT images of the 9 patients with cervical vertebrae(C1 and C2) metastasis. Treatment plans for Xsight spine tracking and 6 Dimensional skull tracking were established with planning system (Multiplan system Version 4.6, Accuray, US). The results of XST and 6DST for each patient were analyzed with Microsoft Excel 2010. Result : The Maximum offsets of XST for C1 were 0.9 mm in Y(supero-inferior), 0.9 mm in Z(antero-posterior), 0.7 mm in X(left-right) direction, and rotations were and 1.0 degrees roll, 1.0 degrees pitch and 1.2 degrees yaw. The Maximum offsets of 6DST for C1 were 0.7 mm, 0.7 mm, 0.9 mm and $1.0^{\circ}$, $1.0^{\circ}$, $1.2^{\circ}$ for Y, Z, X and Roll, Pitch, Yaw. The Maximum offsets of XST and 6DST for C2 were 0.7 mm, 0.7 mm, 0.8 mm and $0.9^{\circ}$, $1.0^{\circ}$, $1.8^{\circ}$, and 0.9 mm, 0.7 mm, 0.9 mm and $0.9^{\circ}$, $0.9^{\circ}$, $1.0^{\circ}$ for Y, Z, X and Roll, Pitch, Yaw, respectively. Conclusion : XST and 6DST showed identical results for translations and rotations within the tolerance. It is possible to simplify the treatment time and procedure by using the 6DST. Therefore, 6DST is very useful methode with XST among the various tracking methods in Cyberknife for the patients with C1, C2 vertebral metastasis.

  • PDF