• Title/Summary/Keyword: Blood flow velocimeter

Search Result 2, Processing Time 0.016 seconds

Design and Evaluation of Blood flow Measurement Using Self-mixing type Semiconductor Laser (자기혼합형 반도체 레이저를 이용한 혈류측정 시스템 설계 및 평가)

  • Kim, Duck-Young;Lee, Jin;Kim, Se-Dong;Ko, Han-Woo;Kim, Sung-Hwan
    • Journal of Biomedical Engineering Research
    • /
    • v.17 no.4
    • /
    • pp.499-506
    • /
    • 1996
  • Blood flow velocimeter is an essential device to measure the blood flow in skin tissue. In this study, we developed a high-speed LDV(laser Doppler Velocimeter) that has real time processing capability using a DSP(digital signal processing) chip and is able to continuously measure information about blood-flow based on a noninvasive method using self-mixing type laser diode. This LDV system has a simpler structure than any other typical blood flow velocimeter and is composed of new self-mixing probe, stabilizer circuits DSP board, and interf'ace boule We measured velocity of speaker-unit by operational frequencies to identify Doppler effect of this system, performed clinical experiment on bare finger tip and compared it with a commercial euipment BPM403A(USA).

  • PDF

Construction of Ultrasound Pulsed Doppler Velocimeter (펄스 초음파 도플러 속도계의 제작에 관한 연구)

  • 현석봉;김수용;이재수
    • Progress in Medical Physics
    • /
    • v.5 no.1
    • /
    • pp.25-39
    • /
    • 1994
  • To measure the velocity of heart wall and local flow transctaneously in blood vessels, we have developed a single channel 3.1 MHz pulsed ultrasonic Doppler velocity meter. Ultrasound pluse width and repetition frequency (PRF) used in the velocity meter is 1 ${\mu}$sec 6kHz reapectively, and the Doppler shift of the backscattered echo signal is sensed in a phase detector by coherent demodulation method. From the output of the phase detector, the Doppler signal corresponding to the mean velocity of acoustic wave scatterers over a small region is obtained by using a range gate, sample holder and band-pass filter. Mean frequency of Doppler signal is estimated by zero-crossing counter and the instantaneous velocity of scatters is displayed as a function of time. It is possible to estimate velocity profile, volume flow and flow acceleration of vessels in man if the number of channels and range resolution in increased.

  • PDF