• Title/Summary/Keyword: Average Grandular Dose

Search Result 1, Processing Time 0.014 seconds

Denoising of Digital Mammography Images Using Wavelet Transform (웨이블릿을 이용한 디지털유방영상의 노이즈 제거)

  • Choi, Seokyoon;Ko, Seongjin;Kang, Sesik
    • Journal of the Korean Society of Radiology
    • /
    • v.7 no.3
    • /
    • pp.181-189
    • /
    • 2013
  • The optimum exposure parameters are found when examined using the automatic mode in FFDM. improve the image quality by applying denoising algorithm and propose methods to reduce AGD(Average Grandular Dose) a patient can receive. For the experiment, Nuclear Associates Model 18-222 phantom was the used, and the entrance dose and AGD were measured. And then, Signal, Noise, SNR and FOM(Figure of Merit) were measured, compared and analyzed image denoising before and after. As the experiment result, first, SNR was the highest at Mo/Mo 23kVp and W/Rh 35kvp was the lowest for the average glandular dose. It showed to use 28kVp of W/Rh to be the best through the result of FOM. SNR was the highest at Mo/Mo 23kVp(image denoising), and it showed to W/Rh and 28kVp to be the best in the FOM result which AGD was considered at the same time. By the image denoising, it is possible to reduce noise while maintain important information in the image.