• Title/Summary/Keyword: Auto Blood Typing Analyzer

Search Result 1, Processing Time 0.015 seconds

Evaluation for the Usefulness of Automated Blood Typing Analyzer (혈액은행 자동화 장비 도입의 유용성 평가)

  • Kim, Ha-na;Kim, Hee-Bum;Park, Hyun-Sang;Lee, Hyun-Im;Hong, Myung-Kook;Shin, Gyoung-Sook;Suh, In-Bum
    • The Journal of the Korea Contents Association
    • /
    • v.19 no.6
    • /
    • pp.565-574
    • /
    • 2019
  • In this study, we evaluated the usefulness of an automatic blood typing analyzer using QWALYS-2 up (Diagast, Loos Cedex, France). During a month( 01OCT2013 - 31OCT2013) we performed 1,636 tests for ABO & RhD blood typing, 1,374 tests for antibody screen & identification tests and compared the results by automatic blood type analyzer with previous manual methods and column agglutination tests. And we analyzed the economic performance by comparison the test unit price between automatic blood type analyzer and manual methods. In ABO & RhD blood typing tests, there were complete concordances between manual and automated blood typing analyzer for 200 clinical samples. In Antibody screen tests, the concordance rate between manual and automated blood typing analyzer was 98.5% and more strong reaction in automated blood typing analyzer than manual methods. Therefore, the introduction of an automated blood typing analyzer, reagents costs were increased but labor costs were decreased. Considering the importance of transfusion safety and economic advantages, the introduction of an automated blood typing analyzer was very useful.