Feasibility of the AtriClip Pro Left Atrium Appendage Elimination Device via the Transverse Sinus in Minimally Invasive Mitral Valve Surgery

  • Shirasaka, Tomonori;Kunioka, Shingo;Narita, Masahiko;Ushioda, Ryohei;Shibagaki, Keisuke;Kikuchi, Yuta;Wakabayashi, Naohiro;Ishikawa, Natsuya;Kamiya, Hiroyuki
    • Journal of Chest Surgery
    • /
    • v.54 no.5
    • /
    • pp.383-388
    • /
    • 2021
  • Background: Achieving external access to and manual occlusion of the left atrial appendage (LAA) during minimally invasive mitral valve surgery (MIMVS) through a small right thoracotomy is difficult. Occlusion of the LAA using an epicardial closure device seems quite useful compared to other surgical techniques. Methods: Fourteen patients with atrial fibrillation underwent MIMVS with concomitant surgical occlusion of the LAA using double-layered endocardial closure stitches (n=6, endocardial suture group) or the AtriClip Pro closure device (n=8, AtriClip group) at our institution. The primary safety endpoint was any device-related adverse event, and the primary efficacy endpoint was successful complete occlusion of blood flow into the LAA as assessed by transthoracic echocardiography at hospital discharge. The primary efficacy endpoint for stroke reduction was the occurrence of ischemic or hemorrhagic neurologic events. Results: All patients underwent LAA occlusion as scheduled. The cardiopulmonary bypass and aortic cross-clamp times in the endocardial suture group and the AtriClip group were 202±39 and 128±41 minutes, and 213±53 and 136±44 minutes, respectively (p=0.68, p=0.73). No patients in either group experienced any device-related serious adverse events, incomplete LAA occlusion, early postoperative stroke, or neurologic complication. Conclusion: Epicardial LAA occlusion using the AtriClip Pro during MIMVS in patients with mitral valve disease and atrial fibrillation is a simple, safe, and effective adjunctive procedure.