• Title/Summary/Keyword: Artificial valve

Search Result 130, Processing Time 0.025 seconds

Nonlinear Analysis of the Monoleaflet Polymer Valve According to Shape of Supporting Members (지지대 형상에 따른 단엽식 고분자 판막의 비선형 해석)

  • 한근조;안성찬;심재준;김성윤
    • Proceedings of the Korean Society of Precision Engineering Conference
    • /
    • 2001.04a
    • /
    • pp.748-751
    • /
    • 2001
  • Monoleaflet polymer artificial heart valve was known to show remarkable improvement in antithrombosis and pressure drop compared with other type of artificial valve. In this investigation monoleaflet the vertical and horizontal deflection pattern of polymer heart valve with three types of supporting members straight member, and two curved members was analysed using the large deformation nonlinear finite element method.

  • PDF

Design of the monoleaflet polymer valve to minimize stress and displacement (응력 및 변위를 최소화하기 위한 단엽식 고분자 판막의 설계)

  • Han, G.J.;Kim, S.H.
    • Proceedings of the KOSOMBE Conference
    • /
    • v.1993 no.11
    • /
    • pp.99-103
    • /
    • 1993
  • A monoleaflet polymer artificial heart valve which showed the remarkable improvement in pressure drop compared with other types of artificial valve was designed to decrease the deflection in vertical direction and the displacement or the valve tip in horizontal direction. Stress distribution change was studied as the location of the supporting members or the valve frame changed. And it was found that using the valve tip horizontal displacement the minimum valve thickness could be obtained in order to prevent the gap between the valve tip and the frame wall.

  • PDF

Neural Network Approach to Automated Condition Classification of a Check Valve by Acoustic Emission Signals

  • Lee, Min-Rae;Lee, Joon-Hyun;Song, Bong-Min
    • Journal of the Korean Society for Nondestructive Testing
    • /
    • v.27 no.6
    • /
    • pp.509-519
    • /
    • 2007
  • This paper presents new techniques under development for monitoring the health and vibration of the active components in nuclear power plants, The purpose of this study is to develop an automated system for condition classification of a check valve one of the components being used extensively in a safety system of a nuclear power plant. Acoustic emission testing for a check valve under controlled flow loop conditions was performed to detect and evaluate disc movement for valve failure such as wear and leakage due to foreign object interference in a check valve, It is clearly demonstrated that the evaluation of different types of failure types such as disc wear and check valve leakage were successful by systematically analyzing the characteristics of various AE parameters, It is also shown that the leak size can be determined with an artificial neural network.

Axisymmetric analysis of blood flow for a floating type polymer artificial heart valve (부유식 폴리머 인공심장 밸브의 축대칭 혈류 해석)

  • Seong H. C.;Jung K. S.;Kim K. H.;Ko H. J.;Park C. Y.;Min B. G.;Shim E. B.
    • Proceedings of the KSME Conference
    • /
    • 2002.08a
    • /
    • pp.703-704
    • /
    • 2002
  • The two major problems related to the blood flow in a floating type polymer valve are thrombus formation and hemolysis. It is well known that the shear stress in the fluid and flow separation around the valve are blamed for such disastrous phenomena. In this viewpoint, through study of the flow field around the valve is imperative to improve design of the valve. The aim of this study is to investigate the fluid flow around a floating type polymer valve. The numerical method employed in this study is the finite element software called ADINA. Incompressible viscous flow is assumed for blood using the assumption of Newtonian fluid. In this study, two prominent features of the axisymmetric flow around the floating type polymer valve are observed: jet-like flows observed near the gap between the conduit and the valve, and recirculating flow downstream of the valve. We also provided a detailed description of shear stress field according to the variation of flow conditions. The shear stress in fluid has its maximum value near the gap between the valve and the conduit.

  • PDF

The Development of a New Polymer Valve for Artificial Blood Pumps (인공심장 및 심실보조장치용 고분자 인조판막의 개발)

  • Suh, S.W.;Wetering, J.E.v.d.;Park, Y.J.;Park, S.K.;Kim, I.Y.;Min, B.G.
    • Proceedings of the KOSOMBE Conference
    • /
    • v.1993 no.11
    • /
    • pp.104-106
    • /
    • 1993
  • Some cardio-vascular assist systems need more inexpensive artificial heart valves for short-term use. To meet with this need, we have developed a new polymer valve that is very simple to manufacture and of which its dimensions are easy to change, depending on its application. We have tested the hydrodynamic performance of the new polymer valve using a mock circulatory testing system and studied the flow through the valve using high-speed camera combined with image processing techniques. The results show that this valve is superior in its performances to the other valves (Bjork-Shiley mechanical valve and trileaflet polymer valve) and have no stagnation points. We also have tested the hemolytic potential of the valve. The valve is less hemolytic than the Bjork-shiley mechanical valve finally, we have applied this valve to a left ventricular assist device that we are developing.

  • PDF

A Study of the Acoustical Properties of the Mechanical Heart Valve Using MUSIC (MUSIC을 이용한 기계식 심장 판막의 음향 신호 특성 연구)

  • Yi S. W.;Choi M. J.;Min B. G.
    • Proceedings of the Acoustical Society of Korea Conference
    • /
    • autumn
    • /
    • pp.131-134
    • /
    • 1999
  • This paper considers the acoustical characteristics of the mechanical valve employed in the Korean type Artificial Heart. $Bj\"{o}rk-Shiley$ tilting disc valve was chosen for the study and acoustic measurements were performed for the artificial heart operated in a mock circulation system as well as implanted to an animal as a Bi Ventricular Assist Device (BVAD). In the mock system, three different conditions of the valve were examined which were normal, damaged (torn off), pseudothrombus attached. Microphone measurements for the BVAD were carried out at a regular time interval for 5 days after the implantation operation. Of the recorded acoustic emissions from the artificial heart, click sounds mainly originated from the valves were further analyzed using Multiple Signal Classification (MUSIC) for estimating their spectral properties. It was shown that the spectral peaks below 4 kHz and the optimal order number for MUSIC, equivalent to the number of the spectral component, might be the key parameters which were highly correlated to the physiological states of the valve like the mechanical damage of the valve or the formation of thrombus on the valves.

  • PDF

In vitro study of downstream flow passing through heart valve prostheses (인공 심장판막 하류 유동의 in vitro 연구)

  • Kim Hyoung-Bum;Lanning Craig
    • Journal of the Korean Society of Visualization
    • /
    • v.4 no.1
    • /
    • pp.47-55
    • /
    • 2006
  • The left ventricular filling flow is now considered as an indicator which can be used for early diagnosing of cardiovascular diseases. Because the understanding of left ventricular flow physics is critical for this purpose, the downstream flow characteristics of the artificial heart valve are investigated using particle image velocimetry (PIV) method. In this study, we investigated the wake characteristics of flows passing through three different artificial valves (St.Jude medical bileaflet mechanical valve. Bjork-Shiley monostrut mechanical valve and St.Jude medical Biocor bio valve). The downstream flow field has remarkably altered according to the different valves. SJM MHV has the flow field similar to the pulsating circular jet and BS MHV has oblique pulsating jet. SJM BHV shows the similar flow field of clinical data of normal heart.

  • PDF

Design of a Mechanical Artificial Heart Valve Prosthesis Appliing Design Methodology (설계방법론을 이용한 기계식 인공심장판막의 설계)

  • 천길정;류형태
    • Journal of Biomedical Engineering Research
    • /
    • v.19 no.3
    • /
    • pp.291-296
    • /
    • 1998
  • A new mechanical heart valve prosthesis has been designed appling systematic design methodology. The function of the heart valve was defined, and search for design variation has been carried out according to the functional structure, Optimal model among the various variations was determined in view of the design specificationn. Proto type valve was fabricated and test has been carried out using a mock circulation system. It has been observed that the pressure profile, cardiac output and behavior characteristics are generally satisfactory.

  • PDF

Finite Element Analysis of the Monoleaflet Polymer to Minimize Stress and Displacement (응력 및 변위를 최소화하기 위한 단엽식 고분자 판막의 유한 요소 해석)

  • 한근조;안성찬
    • Journal of Biomedical Engineering Research
    • /
    • v.17 no.1
    • /
    • pp.85-92
    • /
    • 1996
  • A monoleaflet polymer artificial heart valve which showed the remarkable improvement in pressure drop compared with other types of artificial valve was designed to minimize the deflection in vertical direction and the displacement of the valve tip in horizontal direction obtained by using finite element method as the location of the supporting members of the valve frame changed stress distribution change was also studied on each model generated by changing the distance between the frame and supporting members. It was found that by using the valve tip horizontal displacement the minimum valve thickness could be obtained in order to prevent the gap between the valve tip and the frame wall.

  • PDF

Clinical experience of open heart surgery -70 cases- (개심술 70예의 임상경험)

  • 조광현
    • Journal of Chest Surgery
    • /
    • v.19 no.4
    • /
    • pp.644-662
    • /
    • 1986
  • Seventy cases of open heart surgery were performed in the department of Thoracic and Cardiovascular Surgery, Pusan Paik Hospital, Inje College, from Oct. 1985 to Oct. 1986. And the results were summarized as follows. 1. Among the 70 cases, there were 48 cases of congenital heart anomalies and 22 cases of acquired rheumatic valvular heart diseases. Age range of the congenital patients was 7 months to 31 years with the mean age of 10 years, and the acquired patients was 18 to 62 years with the mean age of 40 years. 2. The heart-lung machine used for cardiopulmonary bypass was Sarns 7000, 5-head roller pump, and the number and type of oxygenators were 5 of membrane type and 65 of bubble type. For all cases GIK [glucose-insulin-potassium] solution was used as cardioplegic solution for myocardial protection during operation. 3. Among the 48 congenital anomalies, there were 12 cases of ASD group, 29 of VSD group, 3 of ECD, 3 of TOF and one of PDA + MR, and to all of which the appropriate radical operations were applied. 4. Among the 22 acquired valvular diseases, there were 11 cases of mitral valve diseases [MS; 4, MSr; 3, MRs; 4], 3 cases of aortic valve diseases [AR:1, ARs;1, ASr;1], 4 cases of double valve diseases [MRs+TR; 3, MRs+ARs; 1] and 4 cases of triple valve diseases [MSr+ASr+TR; 3, MSr+Ar+TR; 1]. To all the diseased mitral and aortic valves, artificial valve replacement was applied except one [As], in which valve plication was applied. And to all the diseased tricuspid valve, DeVega annuloplasty was applied. 5. The number of replaced artificial valves were 29 in 25 patients [congenital; 3, acquire; 22]. In MVR, 6 of mechanical valves [St. Jude Medical valve; 6] and 15 of tissue valves [Carpentier-Edward valve; 11, lonescu-Shiley valve; 4] were used. In AVR, 6 of mechanical valves [St. Jude Medical valve; 6] and 2 of tissue valves [Carpentier-Edward valve; 2] were used. 6. Postoperative complications were occurred in 12 cases. Among them 11 cases were recovered with intensive cares, but one patient [VSD + Fistula of Valsalva sinus] was expired with low cardiac out put syndrome.

  • PDF