• Title/Summary/Keyword: Amperometric detennination of hydrogen peroxide

Search Result 1, Processing Time 0.013 seconds

Spinach Root-Tissue Based Amperometric Biosensor for the Determination of Hydrogen Peroxide (시금치 뿌리 조직 바이오센서를 이용한 과산화수소의 정량)

  • Lee, Beom-Gyu;Yoon, Kil-Joong;Kwon, Hyo-Shik
    • Analytical Science and Technology
    • /
    • v.13 no.3
    • /
    • pp.315-322
    • /
    • 2000
  • The response characteristics of the bioelectrode developed by the co-immobilization of spinach root tissue and ferrocene in a carbon paste matrix for the amperometric determination of hydrogen peroxide were evaluated. In the range of electrode potential examined (-0.3~0.0V vs. Ag/AgCl). the response time was relatively short ($t_{95%}=11.8$ sec) and it responded in the wide range of pH. Also, its detection limit was $2.25{\times}10^{-6}M$ (S/N=3) and a relative standard deviation of the measurements which were repeated 15 times using $1.0{\times}10^{-3}M$ hydrogen peroxide was 1.87%. The bioelectrode sensitivity decreased to 40% of the original value in 19 days of continuous use.

  • PDF