A Study on the Detection Ability of Minute Lesions in X-ray Using the Molybdenum Target (Molybdenum 저지극을 이용한 X-ray의 미세병소 검출능력에 관한 연구)

  • Yang, Da-Rae;Dong, Kyung-Rae;Park, Yong-Soon;Ji, Youn-Sang;Kim, Young-Keun;Kim, Chang-Bok
    • Journal of Radiation Protection and Research
    • /
    • v.35 no.1
    • /
    • pp.43-48
    • /
    • 2010
  • Beam quality is determined according to Xray tube's target material. In a range of between 22 kVp and 28 kVp, molybdenum target generates the characteristics energy between the average 17.9 kVp and 19.5 kVp, which produces the high contrast image of the breast. In this study, we used the Mo/Mo combination breast device and ALVIM TRM phantom and measured the detection ability of the minute lesion in the breast imaging throughout analyzing ROC curves. Assuming that an average subject thickness of the breast is 40 mm, the detection ability was not dependent on the kVp changes in a while dependent on both the mAs and thickness change. We can assure that it is not needed to increase the kVp for the imaging of breast which thickness is within the mean range of 40 mm.