• Title/Summary/Keyword: 화학 양론적 교정

Search Result 1, Processing Time 0.02 seconds

Evaluation of Clinical Risk according to Multi-Leaf Collimator Positioning Error in Spinal Radiosurgery (척추 방사선수술 시 다엽콜리메이터 위치 오차의 임상적 위험성 평가)

  • Dong‑Jin Kang;Geon Oh;Young‑Joo Shin;Jin-Kyu Kang;Jae-Yong Jung;Boram Lee
    • Journal of radiological science and technology
    • /
    • v.46 no.6
    • /
    • pp.527-533
    • /
    • 2023
  • The purpose of this study is to evaluate the clinical risk of spinal radiosurgery by calculating the dose difference due to dose calculation algorithm and multi-leaf collimator positioning error. The images acquired by the CT simulator were recalculated by correcting the multi-leaf collimator position in the dose verification program created using MATLAB and applying stoichiometric calibration and Monte Carlo algorithm. With multi-leaf collimator positioning error, the clinical target volume (CTV) showed a dose difference of up to 13% in the dose delivered to the 95% volume, while the gross tumor volume (GTV) showed a dose difference of 9%. The average dose delivered to the total volume showed dose variation from -8.9% to 9% and -10.1% to 10.2% for GTV and CTV, respectively. The maximum dose delivered to the total volume of the spinal cord showed a dose difference from -14.2% to 19.6%, and the dose delivered to the 0.35 ㎤ volume showed a dose difference from -15.5% to 19.4%. In future research, automating the linkage between treatment planning systems and dose verification programs would be useful for spinal radiosurgery.