APMP (Asia Pacific Metrology Programme) Regional Intercomparison Results of Acoustic Calibrators (음향교정기의 아시아.태평양 지역 국제비교 결과)

  • 서상준;서재갑;조문재
    • The Journal of the Acoustical Society of Korea
    • /
    • v.22 no.6
    • /
    • pp.463-471
    • /
    • 2003
  • The results of the APMP (Asia Pacific Metrology Programme) regional intercomparison of acoustic calibrators were reviewed and analyzed. The artefacts used in intercomparison are a sound level calibrator and a pistonphone. The microphones used to measure the output pressure level are 1" and 1/2" standard microphones (LS1P, LS2P) as well as 1" and 1/2" reference microphones (WS1P/F, WS2P/F). The results obtained using standard microphones are satisfactory, but those obtained by the reference microphones, even though E/sub n/ values are within ±1.0, showed great deviations. Such results had come from the inaccurate calibration of reference microphones. By using the correct calibration results which were obtained by the recently established international standards, the new results were very similar to those of the foreign standard institutes.