Monte Carlo Simulation of Transmission-Type X-ray Tube with Dual-Structured Target (이중 적층 구조 표적을 갖는 투과형 엑스선관의 몬테카를로 전산모사)

  • Kwon Su, Chon
    • Journal of the Korean Society of Radiology
    • /
    • v.17 no.1
    • /
    • pp.107-114
    • /
    • 2023
  • X-ray fluorescence analysis has been widely used in the field of science and industry because it gives information about elements and their concentrations without destruction of samples. To increase analysis accuracy of fluorescence generated by photons of the transmission-type X-ray tube for mixture and compound samples would be recommend to have strong energy near 10 keV and 20 keV simultaneously. Tungsten of 9.65 keV and molybdenum of 17.48 keV were considered as targets with dual deposition structure for obtaining two strong characteristic X-rays, and the transmission-type X-ray tube was analyzed using Geant4 Monte Carlo simulation. The W-Mo structure resulted in strong characteristic X-ray near 10 keV and 20 keV simultaneously. A structure with Mo-W multilayers of 5 ㎛ thick also gave optimal spectrum. Various material combination and thickness optimization for the dual-structured target can give X-ray spectrum with strong characteristic X-ray of specific energies.