• Title/Summary/Keyword: 킬로볼트씨티

Search Result 1, Processing Time 0.014 seconds

Accuracy and radiation exposure from image-guidance in Tomotherapy Hi-Art System (토모테라피 Hi-Art System의 영상유도 정확성 평가와 환자 피폭에 관한 연구)

  • Jang, Jae-Uk;Lim, Hyun-Soo;Han, Man-Seok;Kim, Yong-Kyun;Jeon, Min-Cheol
    • Journal of Digital Convergence
    • /
    • v.11 no.10
    • /
    • pp.577-584
    • /
    • 2013
  • IGRT(Image Guided Radiation Therapy) in radiation therapy is a very useful technique in order to increase setup of patient and position reproducibility. Tomotherapy can increase accuracy of setup to take IGRT by MVCT, but it be for verified accuracy of Image guided, and MVCT occurs the exposure of patient. Through this study, IGRT accuracy of Tomotherapy is very accurate within 1.0mm. When MVCT using Tomotherapy phantom for QA, QC be taken, exposure dose is Fine(2mm Slice thickness) 3cGy, Normal(4mm Slice thickness) 1.5cGy, Corse(6mmSlice thickness) 1.0cGy. Measurement value of spatial resolution using AAPM CT performance phantom did't cause a big difference. As a result, ability of IGRT in Tomotherapy is very accurate. While obtaining image for IGRT, we should minimize expose range because patient's be exposed to radiation. We should make an effort to do accurate radiation therapy to minimize exposure of patient by selecting the appropriate thickness of MVCT depending on patient's body and treat area.