• 제목/요약/키워드: 코리올리스 질량유량계

검색결과 1건 처리시간 0.015초

코리올리스 질량유량계의 유량측정에 영향을 미치는 인자에 관한 실험적 연구 (An Experimental Study on the Influential Factors of Flow Measurement with Coriolis Mass Flowmeter)

  • 임기원;이완규
    • 대한기계학회논문집B
    • /
    • 제27권12호
    • /
    • pp.1699-1707
    • /
    • 2003
  • Coriolis mass flowmeter(CMF), which can measure the mass flow directly, is getting rapid attention for the industrial and custody transfer purpose. In order to study the characteristics and the applicability of CMF, it is tested with the national flow standard system. Two types of sensing tube, U-type and straight type, are employed in the test. Water, spindle oil and viscosity Standard Reference Material whose viscosities are 1, 20 and, 67 $\textrm{mm}^2$/s, respectively, are studied. It is shown that the linearity of CMF is getting deteriorated as the fluid viscosity increases, which is due to the zero drift and the relaxation time of the fluid. To test its applicability in the case of high pressured gas, it is calibrated using compressed air, It shows 1∼l.6 % deviations compared to the calibration results using water. It concludes that the fluid velocity in CMF should be lower than the sonic velocity. In addition, the effects of the vibration from the pipeline and pump on CMF as well as the long term stability are studied.